5.9 Alerts
Beeper sounds once every 30 seconds for 1 second,
red LED blinks every 3 seconds - measured values
exceed selected alarm settings (not with default
settings). LogConnect can set alarm levels via
Software.
If an alarm level has occurred an X will be displayed on the display bottom. On the corresponding PDF
report the alarm status will be indicated, too. If the measurement channel is displayed where an alarm
occurred the X on the right bottom of the display is blinking.
The X disappears when the instrument has been restarted for recording!
Red LED blinks once every 4 seconds. Replace
battery.
Blinks twice or more each 4 sconds. Hardware fault!
5.8.1 Set Date and Time after battery replacement via button
After battery replacement or power, the instrument automatically interrupts changes in configuration
mode to set date, time and interval. If no button will be pressed for 20 seconds the unit proceeds with
the last date and time in memory:
Press N =
No change of date and
time, or Press Y = Yes for
changing date and time
Press Mode button to increase the value, press
Start button to jump to the next value.
After date-time request the Interval (INT) can be
changed.
Press N = No to
abort changes, or
Press Y = Yes to
confirm changes
→
→
→
→
...
*
Download free LogConnect software:
www.dostmann-electronic.de/home.html -> Downloads -> Software / <your logger> / Software /
LogConnect_XXX.zip (XXX choose latest version)
6. Technical data
Temperature
Measuring range:
-
+ 40 ° C
Resolution:
0.1 ° C
Pressure
only LOG220 / LOG220 E
Pressure
only LOG220 / LOG220 E
Pressure
only LOG220 / LOG220 E
Measuring range:
300-1100hPa
Resolution:
0.1 hPa
Memory:
LOG200:
150,000 measurements (Temp. + Exercise)
LOG210:
100,000 data set (Temp., Humidity, dewpoint and motion)
LOG220:
60,000 data sets (Temp., Humidity, pressure, dewpoint and
motion)
LOG200 TC:
60,000 measurements (Temp. + 2 external thermocouples T1, T2
+ motion)
LOG210 TC:
60,000 measurements (Temp., Humidity, dewpoint, + 2 external
thermocouples T1, T2 and motion)
LOG200 E:
100,000 measurements (Temp. + Exercise)
LOG220 E:
60,000 measurements (Temp. Humidity, pressure, CO
2
60,000 measurements (Temp. Humidity, pressure, CO
2
dew point and motion)
Interface:
USB (integral)
Dimensions:
91.5 x 42 x 20 mm
Weight:
Approx. 83 g, Incl. Wall holder
Battery:
1 x 3.6 Volt Lithium ½ AA
Note!
After replacing the battery please check the correct time and date of the internal clock. For
Note!
After replacing the battery please check the correct time and date of the internal clock. For
setting the time see next chapter or 5.2.2.1 LogConnect configuration software.
6.1 Risk of injury:
•
Keep these devices and the batteries out of reach of children.
•
Batteries contain harmful acids and may be hazardous if swallowed. If a battery is swallowed, this
can lead to serious internal burns and death within two hours. If you suspect a battery might have
been swallowed or otherwise caught in the body, seek medical help immediately.
Rel. Humidity
only LOG210 / LOG220 / LOG210 TC /
Rel. Humidity
only LOG210 / LOG220 / LOG210 TC /
Rel. Humidity
only LOG210 / LOG220 / LOG210 TC /
LOG220 E
Measuring range:
0..100% rF
Resolution:
0.1% rF
CO
2
only LOG220 E
CO
2
only LOG220 E
CO
2
only LOG220 E
Measuring range:
400-10000 ppm
Resolution:
1 ppm