14
User Guide
3A/m
Power Frequency
50/60Hz
Magnetic Field
IEC 61000-4-8
3A/m
Power frequency magnetic
fields should be that of a
typical commercial or hospi-
tal environment.
Voltage Dips/
Dropout
IEC 61000-4-11
>95% Dip for
0.5 Cycles
60% Dip for 5 Cycles
30% Dip for 25 Cycles
N/A
N/A
The
display is intended for use in the electromagnetic environment specified below.
The customer or user of the
display should ensure
that it is used in such an
environment.
Guidance and Manufacturer’s Declaration - Immunity
N/A
Surge
IEC 61000-4-5
±1kV Differential
±2kV Common
N/A
EFT
IEC 61000-4-4
N/A
N/A
±2kV Mains
±1kV I/Os
ESD
IEC 61000-4-2
±6kV Contact
±8kV Air
±6kV Contact
±8kV Air
Floors should be wood,
concrete, or ceramic tile. If
floors are synthetic, the r/h
should be at least 30%.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment - Guidance