EN
User manual - English
59141002
5 / 70
Mach LED 130
V1
2021-05-07
Congratulations on acquiring a new MACH LED 130 examination
light.
Please read these Instructions for use very carefully.
Instructions for safe use
1.1 Intended user
The Mach LED 130 is a Class I medical product and may only be
operated by trained medical staff.
1.2 Information and obligation of the user to check the product
Pay attention to the instructions for use when handling the lamp.
These Instructions for use are part of the product and must therefore
be stored in a place close to the product in order for the safety in-
structions and important information to be consulted at any time.
Make sure that the lamp is in satisfactory working order before every
use. If there is obvious damage, unusual operating conditions, etc.,
the lamp must not be used.
1.3 Availability of the instructions for use
These instructions for use and a detailed manual with further tips on
how to use the light can be found online at the following link:
https://dr-mach.de/login/mach-led-130.html
1.4 Intended use / contra-indications
The Mach LED 130 examination light is designed to illuminate an ex-
amination site in medical facilities (e.g. in a laboratory, in hospitals or
doctor's practice) with focused, low-glare, shadow-free light. It ena-
bles the user to perform a diagnosis or carry out medical interven-
tions. The Mach LED 130 light is an operating light that is not fail-safe
when used as a single light. This device is not designed to operate in
areas which are subject to explosions.
Continuous illumination of the open human eye should be avoided
when illuminating the face area.
1.5 Technical data
Class of protection
I
IP protection class
IP 42
Input voltage
(power supply)
100-240 V AC, 50/60 Hz
Input voltage
(light body)
24 V DC
Power consumption
28 W
Current
1.2 A
Examination time
Continuous operation possible
Expected life
1
10 years
1
At the end of the expected (designed) service life, the lamp must be ser-
viced more frequently for safe operation (see the manual for details).
1.6 Lighting technical data
Mach LED
130F
Mach LED
130
Mach LED
130 Plus
Central light in-
tensity (dis-
tance 1 m)
70,000 lux
60,000 lux
100,000 lux
Light field dia-
meter d10
150 mm
165 mm
143 mm
Light field dia-
meter d50
86 mm
91 mm
82 mm
Residual light
intensity (one
shadower)
0%
0%
0%
Residual light
intensity (two
shadowers)
41%
51%
48%
Residual light
intensity
(normed tube)
100%
100%
100%
Residual light
density
(normed tube,
one shadower)
0%
0%
0%
Residual light
density
(normed tube,
two shadow-
ers)
42%
51%
49%
Depth of illumi-
nation (20%)
1750 mm
1750 mm
1750 mm
Depth of illumi-
nation (60%)
655 mm
1016 mm
789 mm
Radiation in-
tensity in the
field
(distance 1 m)
250 W/m
2
217 W/m
2
357 W/m
2
Max. radiation
intensity in the
field
(distance 0.85
m)
288 W/m
2
258 W/m
2
439 W/m
2
For a complete overview of technical and lighting specifications, refer
to the manual.
1.7 Installation/Maintenance/Repair
The lamp may only be installed, maintained or repaired by the man-
ufacturer or by specially trained staff. Maintenance must be carried
out at least every two years
1.8 Environmental conditions for operation
Ambient temperature:
+5°C to +40°C
Relative air humidity
30% to 75% RH
Air pressure
700 hPa to 1060 hPa
1.9 Reporting obligations
All serious incidents which have occurred in connection with the prod-
uct must be reported to the manufacturer and the competent author-
ity.
Images on the device
This symbol tells the user to observe the
instructions for use.
Serial number of the product
Part number of the product
Address of the manufacturer
Date of manufacture and country of manu-
facture