background image

 

 

display, the batteries should be replaced immediately.   

10. Cleaning and Disinfecting   

Use an alcohol swab or cotton tissue moistened with alcohol (70% Isopropyl) 

to clean the thermometer casing and the measuring probe. Ensure that no liquid 

enters the interior of the thermometer. Never use abrasive cleaning agents, thin-

ners or benzene for cleaning and never immerse the instrument in water or other 

cleaning liquids. Take care not to scratch the surface of the probe lens and the 

display.   

11. Battery Replacement   

This instrument is supplied with 2 new, long-lives 1.5V, size AAA batteries. 

Batteries need replacing when this icon « »

24

  is the only symbol shown on the 

display. Use a screwdriver to open the battery cover. Replace the batteries – 

ensure correct polarity as shown by the symbols in the compartment. 

Batteries and electronic instruments must be disposed of in accordance 

with the locally applicable regulations, not with domestic waste.   

 

 

12. Technical Specifications 

Type: Non-Contact Thermometer FR200 

Measurement Range: Body Mode: 34.0-42.2  °C (93.2-108.0  °F) 

Object Mode: 0-100.0  °C (32-212.0  °F) 

Resolution: 0.1  °C /  °F 

Measurement Laboratory:  ±0.2  °C, 36.0~39.0  °C 

Accuracy:(±0.4  °F, 96.8~102.2  °F)   

Display: Liquid Crystal Display, 4 digits plus special 

Acoustic: The unit is turned ON and ready for the measurement: 1 short beep Complete the 

measurement:1 long beep (1 sec.) if the reading is less than 37.5  °C (99.5  °F), 10 short 

«beep», if the reading is equal to or greater System error or malfunction:3 short «bi» sounds 

Memory: 

Auto-Display the last measured

 

temperature30 readings recall in the Memory 

Mode 

Backlight: 

The display light will be GREEN for 4 seconds, when the unit is turned ON. The 

display light will be GREEN for 5 seconds, when a measurement is completed with reading 

equal to or lower than 37.5  °C(99.5  °F).The display light will be RED for 5 seconds, when a 

measurement is completed with a reading higher than 37.5  °C(99.5  °F). 

Operating: Body Mode: 16-40.0  °C (60.8-104.0  °F) 

Temperature: Object Mode: 5-40.0  °C (41-104.0  °F) icons   
Storage:-20  °C to +50  °C (-4  °F to 122  °F) 
Temperature:    15-95 % relative 
maximum humidity   
Automatic:Approx. 1 minute after last measurement   

Summary of Contents for FR200

Page 1: ...1 Infrared Forehead Thermometer User Manual ...

Page 2: ...ter is a high quality product incorporating the latest technology and tested in accordance with international standards With its unique technology this thermometer can provide a stable heat interference free reading with each measurement The instrument performs a self test every time it is switched on to always guarantee the specified accuracy of measurements This Thermometer is intended for the p...

Page 3: ...s Thermometer measures Temperature 4 Control Displays and Symbols 5 Changing between Body and Object Mode 6 Directions for Use 7 Changing between Celsius and Fahrenheit 8 How to recall 30 readings in Memory Mode 9 Error Messages 10 Cleaning and Disinfecting 11 Battery Replacement 12 Technical Specifications ...

Page 4: ... advanced infrared sensor ensures that each measurement is accurate and reliable Gentle and Easy to Use Ergonomic design enables simple and easy use of the thermometer It can even be used on a sleeping child thereby causing no disruption It is quick and child friendly Auto Display Memory The last reading is automatically displayed for 2 seconds when the unit is switched ON Multiple Reading Recall ...

Page 5: ...re be unusually low If the measurement result is not consistent with the patient s finding or unusually low repeat the measurement every 15 minutes or double check the result by another core body temperature measurement This instrument comprises sensitive components and must be treated with caution Observe the storage and operating conditions described in the Technical Specifications section Ensur...

Page 6: ... to replace the battery 5 Changing between Body and Object Mode For changing from body to object mode slide the mode switch at the side of the thermometer downwards For switching back to body mode slide the switch up again 6 Directions for Use Measuring in Body Mode 1 Press the ON OFF button The display is activated to show all segments for 2 seconds 2 The last measurement readings will automatica...

Page 7: ... the temperature in the same location since temperature readings may vary according to locations Doctors recommend rectal measurement for newborn infants within the first 6 months as all other measuring methods might lead to ambiguous results If using a non contact thermometer on those infants we recommend to always verify the readings with a rectal measurement It is recommended to take three temp...

Page 8: ... Reading 30 readings in succession Press and release the START button consecutively to recall the last 30 readings in succession Pressing and releasing the START button 3 after the last 30 readings will resume the above sequence from reading 1 9 Error Messages Measured temperature too high Displays H when measured temperature is higher than 42 2 C 108 0 F in body mode or 100 C 212 F in object mode...

Page 9: ...ns Type Non Contact Thermometer FR200 Measurement Range Body Mode 34 0 42 2 C 93 2 108 0 F Object Mode 0 100 0 C 32 212 0 F Resolution 0 1 C F Measurement Laboratory 0 2 C 36 0 39 0 C Accuracy 0 4 F 96 8 102 2 F Display Liquid Crystal Display 4 digits plus special Acoustic The unit is turned ON and ready for the measurement 1 short beep Complete the measurement 1 long beep 1 sec if the reading is ...

Page 10: ...equirements of the Medical Device Directive 93 42 EEC Technical alterations reserved According to the Medical Product User Act a biennial technical inspection is recommended for professional users Please observe the applicable disposal regulations Remark Some electrical and electrical equipments forbid to abandon and disposal at will Inapplicable baby Attention consult accompanying documents TUV N...