QUICK GUIDE BABYLOG® VN500
22 |
Flow monitoring
Calibrating the neonatal flow sensor
Manual calibration of the neonatal flow sensor is necessary:
− During the device check and before use
– At least once a day
– After replacing the neonatal flow sensor
– After medication nebulization
Recalibration is not necessary if the neonatal flow sensor has been unplugged only briefly.
1.
1. Port for the neonatal flow sensor
Neonatal flow sensor