3
Setup and Installation
3-8
Narkomed GS Setup and Installation Manual
Draeger Medical, Inc.
5. Attach a 22 mm breathing hose betwen the 22 mm hose terminal on
the inspiratory valve marked INSPIRATION, and one side of the
Y-piece.
WARNING:
To ensure proper gas flow direction during inspiration and
expiration, the disks in the inspiratory and expiratory valves
must move freely, without sticking.
6. Attach a 22 mm breathing hose between the other side of the Y-piece
and the expiratory hose terminal on the ultrasonic flow sensor.
7. Attach the breathing bag to the swivel bag mount 22 mm terminal
marked BREATHING BAG.
22MM
BREATHING
HOSES
(P/N 9995132)
INSPIRATORY
VALVE W/22MM
BREATHING
HOSE TERMINAL
ABSORBER
Y-PIECE
ABSORBER
MOUNTING
STUD
SWIVEL BAG MOUNT W/22MM
BREATHING BAG TERMINAL
19MM SCAVENGER
HOSE
(P/N 9995230)
APL VALVE W/19MM
SCAVENGER HOSE
TERMINAL
FRESHGAS
LOCKING DEVICE
W/15MM FEMALE
FITTING
FRESHGAS HOSE
W/15MM MALE
FITTING
BREATHING BAG
MANUAL/AUTOMATIC
SELECTOR VALVE
W/22MM HOSE TERMINAL
ABSORBER POLE
EXPIRATORY
VALVE
SU14718B
SET SCREW
(P/N HW10002)
ULTRASONIC
FLOW SENSOR
W/22MM BREATHING
HOSE TERMINAL
TO SCAVENGER
Summary of Contents for Narkomed GS
Page 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 10: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 40: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 46: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 50: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 52: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 90: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 91: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...