Setup and Installation
3
Draeger Medical, Inc.
Narkomed GS Setup and Installation Manual
3-9
WARNING:
The breathing bag acts as a pressure limiting device during
manually assisted and spontaneous ventilation. Therefore,
breathing bags used with the Absorber system must comply
with the pressure/volume requirements of current ANSI
standards. A bag that has been stretched may have
drastically altered compliance characteristics, and altered
conductivity in the case of conductive bags.
8. Connect the 19 mm scavenger hose between the 19 mm terminal
marked SCAVENGER HOSE on the rear of the APL valve and the
19 mm terminal marked SCAVENGER HOSE on the scavenger.
ULTRASONIC
FLOW SENSOR
OXYGEN
SENSOR
SU98101B
VENTILATOR BELLOWS
W/22MM VENTILATOR
HOSE TERMINAL
19MM
SCAVENGER
HOSE
PRESSURE
SENSING PORT
QUICK-CONNECT
FITTING
VENTILATOR
22MM
BREATHING
HOSE
ABSORBER POLE
TO SCAVENGER
APL
VALVE
Summary of Contents for Narkomed GS
Page 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 10: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 40: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 46: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 50: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 52: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 90: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 91: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...