Setup and Installation
3
Draeger Medical, Inc.
Narkomed GS Setup and Installation Manual
3-13
Installing the
Oxygen
Sensor
1. Remove the inspiratory valve dome plug from the inspiratory valve
dome.
2. Insert the sensor cord connector into the OXYGEN SENSOR interface
underneath the rear panel of the ventilator box.
3. Remove any protective covering from the sensor housing.
4. Perform an oxygen sensor calibration as described in “Operation -
Oxygen Monitoring” in the
Narkomed GS Operator's Manual
.
5. Insert the sensor assembly into the inspiratory valve dome by pressing
it into place.
SU00601
SENSOR
HOUSING
INSPIRATORY
VALVE DOME
INSPIRATORY
VALVE DOME
PLUG
(P/N 4106387)
SENSOR
CORD
CONNECTOR
2
VOLUME
SENSOR
OXYGEN
SENSOR
BREATHING
PRESSURE
AUTO/BAG
SELECTOR
REAR PANEL OF
VENTILATOR BOX