Precautions During Maintenance
WARNING !
Never operate the infant warmer system if it
has suffered physical damage or does not
seem to operate properly. In this case, always
refer servicing to properly trained and factory
authorized service personnel.
WARNING !
Always disconnect power supply before
servicing.
WARNING !
Maintenance work on the Babytherm 8004/
8010 may be performed by trained and factory
authorized staff only.
WARNING !
To avoid any risk of infection, clean and
disinfect bassinet and accessories before any
maintenance according to established
hospital procedures – this applies also when
returning units or parts for repair.
Precautions During Care
CAUTION !
Only use detergents/disinfectants with a
ph value of 7 to 9 for phototherapy lenses.
CAUTION !
Do not launder mattresses in an automatic
washing machine. Do not autoclave mattresses.
CAUTION !
Do not expose Babytherm 8004/8010 to high
levels of ultraviolet radiation. This may cause
cracks in the PMMA (acrylic glass) parts.
CAUTION !
Certain components of the bassinet consist of
materials that are sensitive to certain organic sol-
vents sometimes used for cleaning and disinfect-
ing (e.g., alcohols, phenols, halogen releasing
compounds, oxygen releasing compounds,
strong organic acids, etc.). Exposure to such
substances may cause damage that is not always
immediately recognized. Sterilization of the
bassinet itself with ethylene oxide (EtO) is also
not recommended.
CAUTION !
Do not fold gel mattress!
WARNING !
Always disconnect power supply before
cleaning and disinfecting.
WARNING !
Always follow accepted hospital procedures
for handling equipment contaminated with
body fluids.
Important Safety Information
Precautions During Care
Precautions During Maintenance
11
Operating Instructions Babytherm 8004/8010, 2. ed.