Switching on and activating self test
●
Press On/Off button, until it engages = ON.
Running a self test, Babytherm will now check
important functions.
1
The green »Power On«-LED is lit.
2
After approximately 2 seconds, Babytherm displays
the radiant heater mode:
»man.«
– no skin temperature probe connected
or
»Skin«
– at least the yellow ("tummy") skin tempera-
ture probe is connected.
The respective LED is blinking.
3
The preconfigured set values are displayed.
4
If text message
Battery discharged
is showing and
5
yellow LED is lit, the rechargeable battery for power
failure alarm is being recharged. After approximately
30 minutes, the battery will be sufficiently recharged.
5
The yellow LED will be off.
6
If the red
»Inop.«
LED is lit
= device malfunction
,
see "Troubleshooting", page 53.
Testing LEDs, displays and audible alarm
7
Press
k
key.
8
Display shows message:
All displays on, horn on
All display elements remain on for about 2 seconds, all
LEDs are lit; the digital displays show
88.8
simultaneously with a continuous audible alarm from
the enunciator.
8
Display shows message:
All displays off, soft alarm on
All display elements remain dark for about 2 seconds,
simultaneously with an intermittent audible alarm.
This functional test may be performed also during
operation.
●
Check at least once a day.
Testing Readiness for Operation
Before Each Use
24
Operating Instructions Babytherm 8004/8010, 2. ed.
15 min
Babytherm 8010
1
5
10
<36
°
C
°
C
skin
man.
skin
°
C
manual
mattress.
°
C
8
7
15 min
Babytherm 8010
1
5
10
<36
°
C
°
C
skin
man.
skin
°
C
manual
mattress.
°
C
2
3
3
4
5
1
6
%$