Cleaning / Disinfecting / Sterilizing
To prevent any damage, we recommend that only deter-
gents and disinfectants are used that are compatible with
the materials used in the bassinet and its components
and accessories, e.g. surface disinfectants on the basis
of
– aldehydes, or
– quarternary ammonium compounds
for disinfection procedures.
Ensure that all disinfectants are registered with the U.S.
Environmental Protection Agency for use as intended.
Always follow the instruction labels specifically with
respect to prescribed concentrations and the necessary
exposure times.
CAUTION !
Do not expose Babytherm 8004/8010 to high
levels of ultraviolet radiation. This may cause
cracks in the PMMA (Plexiglas
) parts.
CAUTION !
Certain components of the bassinet consist of
materials that are sensitive to certain organic
solvents sometimes used for cleaning and
disinfecting (e.g., alcohols, phenols, halogen
releasing compounds, oxygen releasing
compounds, strong organic acids, etc.).
Exposure to such substances may cause
damage that is not always immediately
recognized. Sterilization of the bassinet itself with
ethylene oxide (EtO) is also not recommended.
Care
The Babytherm 8004/8010 infant warmer system must
be thoroughly cleaned and disinfected after each patient,
or at least once a week. Perform any disinfection
procedures according to established hospital procedures
as well as to the following additional instructions.
For the cleaning and disinfecting of accessories, refer to
their respective operating instructions.
Disassembly
●
Switch off all device(s), pull power plugs), and turn
off all pressurized oxygen and air if used.
●
Remove any ancillary patient care equipment from
unit.
●
Remove mattress from bed.
Always store mattress flat.
●
Take off flexible U-grommets.
CAUTION !
Do not fold gel mattress!
WARNING !
Always disconnect power supply before
cleaning and disinfecting.
WARNING !
Always follow accepted hospital procedures
for handling equipment contaminated with
body fluids.
Care
Disassembly
Cleaning / Disinfecting / Sterilizing
50
Operating Instructions Babytherm 8004/8010, 2. ed.