22
Instructions for use
|
Dräger BG ProAir
Maintenance
The hand wheel for the compressed oxygen cylinder
must be positioned in the depression on the outside of the
carrying frame.
4. Tighten the hand wheel of the pressure reducer to the
compressed oxygen cylinder. Do not use any tools to
tighten.
5. Thread the cylinder strap into the buckle and close the
hook-and-loop fastener. Slide the buckle onto the edge of
the breathing bag so that the lid of the closed-circuit
breathing apparatus can be closed.
5.2.5.9 Closing the closed-circuit breathing apparatus
1. Interlink the hinges and lock them in place.
2. Close the lid and engage the locking mechanisms. A click
can be heard.
5.2.5.10 Fitting the waist belt and shoulder straps
1. Fit the top ends of the shoulder straps. To do so, insert the
connecting elements in the mounts on the carrying frame
and engage. The loops should point towards one another.
2. Fit the bottom ends of the shoulder straps. To do so, insert
the connecting elements in the mounts on the carrying
frame and engage.
3. Fit the waist belt. To do so, position the waist belt plate
with the recess facing upwards on the carrying frame’s
waist belt support and secure with the split pin.
4. Fasten the Connect ECU on the shoulder strap.
5.2.6 Testing the closed-circuit breathing
apparatus
5.2.6.1 Notes for testing
The following test sequence applies for both apparatuses with
constant oxygen dosage and those with on-demand oxygen
dosage. The pre-flush unit does not need to be tested for
apparatuses with constant oxygen dosage. This test is
marked accordingly.
Carry out a visual inspection, as well as a leak and functional
test, after every service. If a fault is detected during the leak
and functional test, the closed-circuit breathing apparatus
must not be used. Appoint qualified personnel to service the
apparatus or send it to Dräger.
If the CO
2
absorber is installed in the closed-circuit breathing
apparatus, the leak and functional test can be performed a
total of six times, whereby testing should not take longer than
15 minutes. After these 6 tests, the CO
2
absorber must be
replaced and must no longer be used.
The tests are described using the Dräger RZ 7000 by way
of example. The sequence of the tests described in these
instructions for use is the optimum sequence with this test
unit. The closed-circuit breathing apparatus can also be
tested with other suitable test units in other sequences.
The tests should not be performed in direct sunlight or in a
cold room as the air pressure in the device will then change
quickly and can result in incorrect test results.
51997
51998
4651
6
Summary of Contents for BG ProAir
Page 2: ......
Page 31: ...Annex Instructions for use Dr ger BG ProAir 31...
Page 32: ......
Page 33: ......