background image

PAS

®

 Micro Series

Airline/Short duration respiratory protection equipment with PAS

®

 ASV

Instructions for Use

i

3359994 (A3-D-P) Page 2 of 3

damage that may affect the operation of the breathing apparatus

include impact, abrasion, cutting, corrosion and discoloration.

Report damage to service personnel and do not use the apparatus

until faults are rectified.

3.1.2

Fitting the cylinder

1. Ensure that the cylinder is fully charged.

2. Place the carrying harness on a clean flat surface.

3. Fully extend the cylinder strap, shoulder straps and waist belt.

4. Check the threads of the cylinder valve port and the pressure

reducer. Ensure that the O-ring seal (Fig 4, Item 1) and the

sintered filter (Fig 4, Item 2) in the reducer are clean and

undamaged.

5. Referring to Fig 5, slide the cylinder through the cylinder strap

and align valve with the reducer. To prevent damage, ensure

that the cylinder remains clear of the hand wheel of the

pressure reducer while inserting the cylinder.

6. Align the cylinder with the centre of the carrying harness and

then screw the hand wheel of the pressure reducer on to the

cylinder valve (see direction arrow in Fig 5). 

Do not fully

tighten

.

7. Referring to Fig 6, take up the slack in the cylinder strap (1) to

tighten around the cylinder and then pull the strap through the

buckle to secure (2). Insert the loose end of strap into the guide

loop (3).

8. Fully tighten the hand wheel hand tight. Do not use tools or

over tighten.

3.1.3

Functional testing

WARNING

If the breathing apparatus fails to meet any of the standards

or parameters described in the functional tests, or if an

immediate leak is evident, there is a system fault. Report

the fault to trained service personnel or contact Dräger. Do

not use the breathing apparatus until the fault condition is

rectified.

1. Positive-pressure systems: press the reset button

(Fig 3, Item 2) to switch off the valve.

2. Open the cylinder valve slowly, but fully, to pressurize the

system and ensure that cylinder is fully charged. The whistle

on the PAS

®

 ASV sounds.

3. Fully close the cylinder valve.

4. Allow the air to vent from the PAS

®

 ASV whistle and observe

the pressure gauge.

5. At approximately 70 bar, cover the whistle to stop the air flow

and then very slowly release the air. The whistle on the

PAS

®

 Micro pressure reducer must commence in the range

110 to 100 bar.

6. Open the cylinder valve slowly, but fully, to pressurize the

system. The whistle on the PAS

®

 ASV sounds.

7. To minimize cylinder air use: 

immediately

 connect the

independent air supply to the male coupling (Fig 1, Item 6),

and if the independent air supply has a shut-off valve, open the

valve. The whistle on the PAS

®

 ASV stops.

8. Check for audible leaks. If there is any leak, investigate and

repair the leak before use (see Section 4). If necessary, use a

soapy solution to locate the leak.

WARNING

Do not direct the air flow on to the face, eyes or skin.

9. Press the front button (Fig 3, Item 3) to activate air flow from

the valve for 3–5 seconds. Unobstructed air will flow from the

outlet of the lung demand valve.

10. Positive-pressure systems: press the reset button

(Fig 3, Item 2) to switch off the valve.

11. Isolate and disconnect the independent air supply, and fully

close the cylinder valve.

12. Wait one minute and then observe the pressure gauge and

reopen the cylinder valve. The gauge must not show an

increase in pressure of more than 10 bar (one radial marking

on the gauge face). Investigate and repair any leak before use

(see Section 4). If necessary, use a soapy solution to locate

the leak.

13. Close the cylinder valve and vent the system as follows:

Positive-pressure systems: cover and seal the lung

demand valve outlet with the palm of the hand. Press the

front button (Fig 3, Item 3) to activate air flow and then lift

the hand to very slowly vent.

Negative-pressure systems: carefully press the front button

(Fig 3, Item 3) to very slowly vent.

14. The PAS

®

 ASV whistle begins to sound at the preset switching

pressure, indicating that the valve has switched from the

independent air supply (AL) to the PAS

®

 Micro  air  cylinder

(BA). The whistle on the pressure reducer will also sound as

the pressure falls. Both whistles will cease when the system is

vented.

15. Positive-pressure systems: press the reset button

(Fig 3, Item 2) to switch off the valve.

3.1.4

Putting on the PAS

®

 Micro (ready position)

See also Fig 1 which shows the PAS

®

 Micro  worn  in  the  ready

position.

1. Fully loosen the shoulder straps and waist belt and put on the

apparatus.

2. Check that the shoulder pads are not twisted and take the

weight of the system on the shoulders by pulling the shoulder

straps. Do not fully tighten at this stage.

3. Close the waist belt buckle and pull the ends of the waist belt

until it fits securely and comfortably.

!

!

4. Pull the shoulder straps until the breathing apparatus rests

securely and comfortably on the hips. Do not over tighten.

5. Check that the face mask port, and the lung demand valve

coupling and O-ring are clean and undamaged.

6. Connect the lung demand valve to the face mask as follows:

Push-in coupling: press into the port of the face mask until

it latches in position. Check the attachment by gently

attempting to pull the coupling apart.

Screw-in coupling: screw into the port of the face mask and

tighten hand tight. When the lung demand valve is fitted to

the face mask, the connector can swivel to allow for head

and body movement of the wearer.

7. Put the neck strap of the face mask over the head, and then

insert the neck strap stud into the hole in the centre strap of the

head harness.

3.1.5

Putting on the face mask

WARNING

Correct fit of the face mask can only be achieved if the

complete mask seal makes contact with skin. Head hair,

facial hair (including beard stubble and sideburns),

earrings, other facial piercings and normal spectacles will

interfere with the mask seal and are not permitted in the

sealing area. Additionally, head hair that could affect the

face mask fit (buns, pony-tails, hairpieces, etc.) is not

permitted.

NOTICE

Refer also to the face mask Instructions for Use.

1. Positive-pressure systems: press the reset button

(Fig 3, Item 2) to switch off the valve.

2. Open the cylinder valve slowly, but fully, to pressurize the

system. The PAS

®

 ASV whistle is supplied from the air cylinder

and therefore sounds immediately.

3. Detach the neck strap stud from the centre strap of the head

harness.

4. Spread the head harness (Fig 7). Place the chin into the chin

cup of the face mask and pull the harness over the head

locating the harness centre plate on back of the head.

5. Referring to Fig 8, tighten both lower (1) and then upper straps

(2) evenly towards the back of the head. If necessary, tighten

the centre strap (3).

6. Breathe normally and carry out the mask function check.

3.1.6

Mask function check

1. Close the cylinder valve and breathe normally to empty the

system of air. When empty, the face mask should hold on to

the face to indicate a positive seal. If a leak is detected,

readjust the head harness and retest.

2. Immediately reopen the cylinder valve and recommence

breathing – exhaled air should flow easily out of the exhalation

valve.

When the function check has been satisfactorily completed,

breathe normally and proceed to the work area.

3.2

During use

The nominal duration (10 or 15 minutes) applies to a PAS

®

 Micro

with a fully charged cylinder, and refers to the usage time until the

whistle on the pressure reducer sounds to warn the wearer that

there is low cylinder pressure. The wearer should be in a safe area

before the whistle on the pressure reducer sounds.

3.2.1

Self-contained use (not supplied by an airline)

WARNING

The available duration starts from the time that the cylinder

valve is opened, and is dependent on the capacity of the

cylinder and the breathing rate of the wearer.
Dräger strongly advise that the front button (Fig 3, Item 3)

is not pressed to deliver supplementary air into the mask

during self-contained use. Using supplementary air would

use air from the cylinder and reduce the operating duration

of the air supply.

Regularly check the reading on the cylinder pressure gauge.

If the whistle on the pressure reducer sounds, there is low

cylinder pressure. Immediately leave the hazardous area by

the shortest and safest route.

The PAS

®

 ASV whistle sounds continuously.

3.2.2

Airline use

WARNING

Dräger recommend that the user carries out a risk

assessment to establish local procedures to be followed in

the event of failure of the independent air supply.
If the PAS

®

 ASV  whistle  sounds  at  any  time  during  an

operation, the PAS

®

 ASV has switched to supply the

wearer from the PAS

®

 Micro air cylinder. If you continue to

work while breathing from the air cylinder, regularly check

the cylinder pressure gauge – ensure that the remaining air

in the air cylinder allows sufficient time to evacuate to a

safe area.
During airline use, at very high work rates the pressure in

the face mask may become negative at peak inhalation

flow.

!

i

i

!

!

NOTICE

The PAS

®

 Micro cylinder valve must remain open during

airline use.

Connect the independent air supply to the male coupling

(Fig 1, Item 6), and if the independent air supply has a shut-off

valve, open the valve. The whistle on the PAS

®

 ASV stops –

this indicates that the wearer is now breathing from the airline.

If additional air is required, briefly press and release the front

button (Fig 3, Item 3) to deliver a single jet of supplementary air

into the mask.

When the task is complete or when advised by a controller

leave the hazardous area:

Disconnect from the independent air supply (see

Section 3.2.1), or

Remain connected and carefully withdraw the airline hose.

Evacuation procedure (independent air supply failure)

If the independent air supply fails, the wearer should proceed in

line with established local procedures (see warnings above).

To evacuate proceed as follows:

a. Disconnect the independent air supply from the

PAS

®

 Micro.

b. Immediately leave the hazardous area by the shortest and

safest route. The PAS

®

 ASV whistle is supplied from the air

cylinder and therefore sounds continuously during the

evacuation.

When in a safe area, remove the lung demand valve from the face

mask if necessary and continue to breathe normally.

3.3

After use

WARNING

Do not remove the equipment until in safe area, clear of

hazard.

CAUTION

Do not drop or throw down equipment as damage could

occur.

1. Loosen the face mask straps.

Positive-pressure systems: as the seal between the mask

and the face is broken, press the reset button (Fig 3,

Item 2) to switch off the valve.

2. Remove the face mask and fully extend all of the straps of the

head harness.

3. If the independent air supply is still connected, isolate and

disconnect the supply.

4. Fully close the cylinder valve.

5. Press the front button (Fig 3, Item 3) to fully vent the system.

6. Remove the lung demand valve from the face mask (Fig 9 –

push-in coupling shown).

7. Open the waist belt buckle, lift the shoulder strap buckles to

loosen the harness, and then remove the equipment.

8. Carry out the after use maintenance tasks in the maintenance

table (see Section 5.1).

4

Troubleshooting

The troubleshooting guide shows fault diagnosis and repair

information applicable to breathing apparatus users. Further

troubleshooting and repair information is available in Instructions

for Use supplied with associated equipment (e.g. face mask and

air cylinder).

Contact service personnel or Dräger when the remedy information

indicates a service task, or if the symptom remains after all remedy

actions have been attempted.

5

Maintenance

5.1

Maintenance table

Service and test the breathing apparatus, including out-of-use

apparatus, in accordance with the maintenance table. Record all

service details and testing. Refer also to the Instructions for Use

for the lung demand valve, face mask and other associated

equipment.

Symptom

Fault

Remedy

High-pressure air 

leak or failed leak test

Loose or dirty cylinder 

connector

Disconnect, clean and 

reconnect the connector 

and retest

Faulty hose or 

component

Substitute user 

replaceable accessories 

and retest

Air leak from 

medium-pressure 

hose connection at 

the pressure reducer 

(safety relief valve)

Faulty O-ring, retainer, 

spring or pressure 

reducer

Service task

High or low medium 

pressure

Pressure reducer fault Service task

Poor sounding 

whistle

Whistle dirty

Clean whistle flute and 

retest

Whistle not 

functioning correctly

Activation mechanism 

fault

Service task

i

i

!

!

Reviews: