Operation
8.6.3 Entering
the
menu
●
To directly enter the
Information
menu:
a. Tap and hold for 1 second in the measuring mode.
●
To enter the
Calibration
menu:
a. Tap and hold for 3 seconds in measuring mode
b. Select
Enter password
c. Enter
Calibration PWD
.
●
To enter all menus:
a. Tap and hold for 3 seconds in measuring mode
b. Select
Enter password
c. Enter
Settings PWD
8.6.4 Displaying
Information
Select the appropriate menu item in the
Information
menu:
M
M
M
Instrument
Warnings
Displays warning messages in plain text and the corre-
sponding number.
If there are several warning messages available, there is an
indicator (e.g. 1/3 = Screen 1 of 3).
Faults
Displays fault messages in plain text and the corresponding
number
If there are several faults, there is an indicator (e.g. 1/3 =
Screen 1 of 3).
Device flag
This function displays warning and fault codes as a table. If
all codes are 00 there are no warning or fault messages
available.
Modules
Displays an overview of the installed hardware modules. To
access detailed information select the appropriate module.
= installed modules
= not installed modules
Sensor
Vitality
1)
The sensor vitality is displayed in %. Changing the sensor
is recommended with a vitality < 25 % by Dräger.
Last cal. date
The calibration date, unit, gas and concentration of the last
calibration are displayed.
Next cal. date
Displays the next calibration due date.
Sensor tempera-
ture
1)
1) Function only available with diagnostic dongle.
The current and the highest sensor temperature is dis-
played.
Data logger
Logger status
Displays the data logger status (active or non-active)
Graph
Displays the history of the past 15 minutes on a time/con-
centration graph.