92
Technical Manual
|
Dräger Polytron
®
8000 Series
Technical data
17.6 Environmental
parameters
See sensor data sheet for sensor specifications and influences.
The values for accessories and spare parts are less strict than or identical to those
of the transmitter. For exact values, refer to the corresponding manual.
17.7 Tightening
torque
Polytron
®
8xx0
Pressure (performance approval) 20.7 to 38.4 in. Hg (700 to 1300 hPa)
Pressure (explosion protection)
23.6 to 32.5 in. Hg (800 to 1100 hPa)
Humidity
0 to 100%RH, non-condensing
Temperature
Polytron
®
8100 EC
1)
1) See specific EC sensor data sheet for limitations to the maximum specification.
-40 to +149 °F (-40 to + 65 °C)
-40 to + 158 °F
2)
(-40 to + 70 °C)
2) Value from Ex approval. For values without this footnote, EX and performance Approval values
are identical.
Polytron
®
8200 CAT
3)
3) Except for relay option and UL approval-relevant installations. For relay option and UL installa-
tions: -40 to + 158°F (-40 to +70°C)
-40 to +176 °F (-40 to +80 °C)
Polytron
®
8310 IR
-40 to +149 °F (-40 to +65 °C)
Polytron
®
87x0 IR
-40 to +170 °F (-40 to +77 °C)
-40 to + 176 °F
(-40 to + 80 °C)
Storage temperature
-4 to +149 °F (-20 to +65 °C)
Storage pressure
26.5 to 32.4 in. Hg (900 to 1100 hPa)
Storage humidity
0 to 100%RH, non-condensing
Instrument threads
Enclosure lid
≥
44 LB IN. /
≥
5 Nm
Sensor
min. 266 LB IN. / min. 30 Nm
Blind plugs
min. 266 LB IN. / min. 30 Nm
Ports (for conduits or cable
glands)
min. 443 LB IN. / min. 50 Nm
Docking Station blind plugs
62 lbs-in / 7 Nm
Docking Station cable glands
70 lbs-in / 8 Nm
Field wiring terminals
Power supply, signals and
relays
4.4 - 7.0 LB IN. / 0.5 - 0.8 Nm
Grounding screws
10.6 LB IN. / 1.2 Nm