Issue 6 – December 2006
12
When the Options board is fitted, the 4-20mA output is on terminals 3 and 4 of the
main board and the TWA relay output is on terminals 20 to 22.
The RS-232 datalogger output is on an RJ11-style 4-position/4-conductor socket.
Data are sent by the controller as ASCII characters
Data rate:
2400 Baud
Data bits:
8
Stop bits:
1
Parity: none
Protocol:
Xon / Xoff
The gas level is and TWA are printed every 1s. For example:
Gas 20.1 TWA 0.1
To reset the TWA value to zero, hold the unit’s Reset pushbutton pressed for 10s, or
hold the remote reset input closed for 10s, until the display shows
tA-0
.
RJ11 socket
RS 232 Sub D – 9 Pole
Pin Function
Pins
Function
1 Rx
3
Tx
2 0
V
5
GND
3 Tx
2
Rx
4 0
V
5
GND
1
nc
6
nc
7
RTS
8
CTS
9
nc
RS-232 data
logger output