Issue 6 – December 2006
21
CONFIGURATION AND CALIBRATION
Set measuring range
Use switch SW1 on the display board to set the display measuring range.
The number of decimal places is fixed according to the measuring range.
For display board with r1
.
00 or r2.00 firmware*, set the range using this table:
SW1
Measuring
range
DPs
A
B
C
D
E
0 - 1
2
OFF OFF OFF OFF OFF
0 - 3
2
ON
OFF OFF OFF OFF
0 - 5
2
OFF
ON
OFF OFF OFF
0 - 10
2
ON
ON
OFF OFF OFF
0 - 25
1
OFF OFF
ON
OFF OFF
0 - 30
1
ON
OFF
ON
OFF OFF
0 - 50
1
OFF
ON ON
OFF OFF
0 - 100
1
ON
ON
ON
OFF OFF
0 - 200
0
OFF OFF OFF
ON
OFF
0 - 250
0
ON
OFF OFF
ON
OFF
0 - 300
0
OFF
ON
OFF
ON
OFF
0 - 500
0
ON
ON
OFF
ON
OFF
0 - 1000
0
OFF OFF
ON ON
OFF
0 - 2000
0
ON
OFF
ON ON
OFF
0 - 2500
0
OFF
ON ON ON
OFF
0 - 3000
0
ON
ON
ON
ON
OFF
0 - 5000
0
OFF OFF OFF OFF
ON
0 - 9999
0
ON
OFF OFF OFF
ON
DPs = decimal places