Issue 6 – December 2006
36
SPECIFICATIONS
Regard-1
4-20
Regard-1
SE Ex
Dimensions
270 mm
u
270 mm
u
90 mm
Mass (without
batteries)
Approx. 2.5 kg
Housing material
ABS – VO
Ingress protection
IP 65
Cable entries
6 No. M20
AC supply voltage
98 to 253 VAC, 50-60 Hz
DC supply voltage
18 to 30 VDC
Power consumption
50 W (without transmitter)
70 W (including sensor)
Mains fuses
500mA (T) HRC.
Battery charge
voltage
- maximum current
27.6
r
0.6 VDC
500
mA
DC output
24
r
1 VDC, 100 mA max.
Battery type (x2)
12 V, 1.2 AH
A1 and A2 outputs
22 VDC at 30 mA
18 VDC at 100 mA
Alarm relays
(A1, A2, A3)
single-pole changeover
max. 250 VAC, 3 A
max. 30 VDC, 2 A
energise on alarm
Fault relay
single-pole changeover
max. 250 VAC, 3 A
max. 30 VDC, 2 A
normally energised (de-energise on fault)
Inhibit relay
single-pole changeover
max. 250 VAC, 3 A
max. 30 VDC, 2 A
energise on alarm
Visual outputs
- with display board
fitted
“AC power on” LED
“DC power on” LED
A1, A2 and A3 alarm LEDs, Fault LED, Inhibit LED
LCD gas level indicator
Transmitter/sensor
input
2- or 3-wire 4-20mA,
24 VDC / 400 mA
3-wire Pellistor
constant current supply
200…350 mA
Storage conditions
–10°C to 60°C
0 to 100% rH, non-condensing
Operating conditions
0°C to 55°C
0 to 100% rH, non-condensing
Warm up time
30 s
Response time
2 s
Approvals and CE
Electromagnetic Compatibility (EMC) Directive
Summary of Contents for REGARD-1
Page 38: ...Issue 6 December 2006 38 EC type examination certificate ...
Page 39: ...Issue 6 December 2006 39 ...
Page 40: ...Issue 6 December 2006 40 ...
Page 41: ...Issue 6 December 2006 41 ...
Page 44: ...Issue 6 December 2006 44 ...
Page 46: ...Issue 6 December 2006 46 ...