Warning! - The charger must not be repaired independently by users
and disassembled by unauthorised persons.
The battery pack should be charged only with the use of the charger
supplied by the manufacturer.
With regard to the user's safety and durability of the device, the
charger must not be used in humid or wet places.
Before using the charger always check if its main elements, including
cables, are not damaged.
Should any damage be identified, immediately disconnect the device from the power source
and replace the damaged part with a new one by contacting an authorised service centre.
Attention! A charger is an auxiliary device used solely for the purpose of battery charging.
It is not an integral part of DRAMI
Ń
SKI Animal profi 2 ultrasound scanner, used during
operation of the device.
Attention!
Connecting the charger to the pack attached to the ultrasound scanner causes automatic
disconnection from the power source and makes examination impossible.
Precautions for dealing with the Li-Ion battery pack
•
Battery packs which form an integral part must not be disassembled.
•
Contacts on the pack casing must not be connected with metal objects.
•
Battery packs must not be thrown into the fire or heated up.
•
Mechanical packs must not be impacted in any way or thrown.
•
If an electrolyte leaks from the pack it, the user must immediately refrain from using
the pack.
•
Do not allow any liquid to permeate into the pack, which may result in a sudden
increase in the battery temperature and cause danger.
•
The battery pack must not be used or stored in high temperature, e.g. inside a car,
directly in the sun, close to heat sources. Failure to comply with these rules may
cause the electrolyte to leak from the battery and damage it or reduce its operating
life.
•
The battery must be charged in the ambient temperature of 0 ° C to 40 °C. Charging
the battery in an ambient temperature different from the one provided above may
cause danger and permanent damage of the pack.
•
If, after a long-term use (about 500 cycles), problems with charging occur, the
battery pack should be replaced.
•
An old battery pack should be recycled in accordance with currently enforced