Quick Reference Guide for PowerXplorer / Power Guide / Power Visa
pg 7
1. Voltage and Current Connections – there are four differential voltage and four differential
current channels, which can be wired to measure a variety of circuit configurations. Use
only Dranetz-BMI voltage leads and current probes for proper operation. Do not exceed
marked ratings.
2. Graphic Color LCD with Touch Screen – the color LCD has an integral touch screen that
is used to select functions for setups, real time meters, and viewing stored data. Either a
clean finger or a PDA-type stylus can be used and the screen should be calibrated with
whichever you are going to use by selecting Preferences; Display Preferences, then
Calibrate Touch screen when first turned on. Clean only with soft cloth.
3. Field Replaceable Battery Pack – the internal UPS feature requires that the internal
batteries be properly charged and maintained. If the battery pack is in need of
replacement, de-energize the connections and power to the unit before opening the
battery door and replacing the pack only with the Dranetz-BMI battery pack PN# BP-
PX5. A spare battery pack may be useful to have.
4. Memory Card slots for data and firmware updates – the Dranetz-BMI supplied Compact
Flash memory cards are used for data storage, as well as to upload new firmware
versions when available. Use only Dranetz-BMI supplied memory cards, as the speed