60
DranTech X
TRA
User’s Guide
Maintenance – Calibration
9
Maintenance and Calibration
Attention!
!
Disconnect the instrument from the measuring circuit before
opening the battery compartment lid or fuse cover in order to re-
place batteries or fuses!
9.1
Displays – Error Messages
9.2
Batteries
Note!
Removing the Batteries During Periods of Non-Use
The integrated quartz movement draws power from the batteries
even when the instrument is switched off. It is advisable to
remove the batteries during long periods of non-use for this rea-
son (e.g. vacation). This prevents excessive depletion of the bat-
tery, which may result in damage under unfavorable conditions.
Note!
Battery Replacement
Stored measurement data is lost when the batteries are replaced.
In order to prevent data loss, it is advisable to backup your data
to a PC with the help of DranWin 10 software before replacing
the batteries. The selected operating parameters remain in mem-
ory, although date and time must be reentered.
Battery
The current battery charge level can be queried in the “
1nfo
” menu:
1nFo
bAtt
:
2.75 V.
Make sure that no battery leakage has occurred before initial start-up,
as well as after long periods of storage. Continue to inspect the batter-
ies for leakage at short, regular intervals.
If battery leakage has occurred, carefully and completely clean the
electrolyte from the instrument with a damp cloth, and replace the bat-
teries before using the instrument.
If the “
” symbol appears at the display, the batteries should be re-
placed as soon as possible. You can continue working with the instru-
ment, but reduced measuring accuracy may result.
The instrument requires two 1.5 V batteries in accordance with
IEC R 6 or IEC LR 6, or two equivalent rechargeable NiCd batteries.
Message
Function
Meaning
FUSE
Current measure-
ment
Blown fuse
In all operating
modes
Battery voltage has fallen below 1.8 V
0L
Measurement
Indicates overflow
MEASURE
SETUP
FUNC
ENTER
Summary of Contents for DranTech Xtra
Page 1: ...DranTech XTRA TRMS Digital Multimeter User s Guide P N M240H 2007 Rev A...
Page 5: ...DranTech XTRA User s Guide 5 Contact Persons...
Page 11: ...DranTech XTRA User s Guide 11 Safety Precautions...
Page 25: ...DranTech XTRA User s Guide 25 Control Functions...
Page 43: ...DranTech XTRA User s Guide 43 V Hz Temperature and A Hz Measurements...
Page 49: ...DranTech XTRA User s Guide 49 Device and Measuring Parameters...