F) How to Shoot the Troubles in the Fluorescence Microscope:

If there are some troubles in operating, please recheck the instrument carefully as the

following describing before connect with us or our representative in your area.

1.

Troubles in electric system:

Troubles

Cause

Remedies

The power supply

box can not be

started

Power cord is not be connected

Connect the box wit the power source

correctly

Fuse is broken

Change it.

The mercury can

not work

The mercury lamp is not

electrified

Connect the lamp house with the

power supply box correctly.

The mercury lamp is broken

Change it

2. Troubles in optical system

Troubles

Cause

Remedies

Nothing or too

dark in the field

after turn on the

mercury lamp.

Flashboard B (6) is not in its position

Push the flashboard B (6) in

correct position

the selector (5) is not in its position

Push the selector (5) in correct

position

The filter system is inconsequence

Select appropriate lens

The image is

blurring and low

contrast

Objective or filter is dirty

Clean it

The filter system is inconsequence

Select appropriate lens

Immersion oil produces the fluorescence

Use special immersion oil

The lamp in base is not turned off

Turn off the lamp

The lenses in collector reflect ambient light

to disturbance the fluorescent observation

Close the field diaphragm in

base or cover the collector

with pressboard

The brightness of

the image is not

even

The nosepiece is not in its position

Turn the nosepiece in correct

position

Flashboard B (6) is not in its position

Push the flashboard B (6) in

correct position

Epi-beam is not in the optical path

Correct it.

The mercury lamp is not centered, or the

condenser is not in focus

Readjust it

G) Maintenance and Care of Microscope:

1. Unpack the microscope carefully to prevent the accessories such as lens from falling down

and damaging.

2. All lens are calibrated, don’t try to dismantle then apart by yourself.

3. Nosepiece and focusing system are advanced and precise in construction. Don’t try to

dismantle them apart by yourself. Please connect with an authorized technician when they are in

trouble.

4. Keep the mechanical parts from dust, and add a few no-corrosiveness lubricating grease into

the sliding sections at regular intervals. Keep the optical elements clean when

P.6

Summary of Contents for DW-2002H

Page 1: ...EPI FLUORESCENT MICROSCOPE SERVICE MANUAL Read it carefully before operating and keep it with the microscope...

Page 2: ...oscope It is not only used for fluorescent microscopy observation but also used for general biological microscopy observation 2 No dismantle the equipment Unless you are a microscopic expert or there...

Page 3: ...the microscope otherwise it will damage the circuit and bulb even lead to insecurity 5 Protecting optical parts Never try to contact directly the optical surface of objectives eyepieces and other opt...

Page 4: ...Microscope P 2 C The Principle Function and Main Technical Parameters of the Fluorescence Microscope P 3 D How to Use the Fluorescence Microscope P 4 E Use Scope of the Fluorescent Filter P 5 F How t...

Page 5: ...let Bar 13 Fluorescent Objective 14 Microscope Body 15 Brightness Adjustor 16 Tension Guide Line 17 Switch 18 Accumulator 19 Power Supply Box 20 Ultraviolet Bar 21 Observation Window 22 Epi fluorescen...

Page 6: ...with soft cloth and gasoline promptly Remove the plastic cover away from the Epi fluorescence device 22 and install it on the microscope body 14 with the screw 11 Usually the lamp house in the Epi fl...

Page 7: ...the power supply box 19 and fix it 5 If the objectives are packed in plastic bottles please take them out and spin them in the nosepiece 6 Install the viewing head 3 on the Epi fluorescence device 22...

Page 8: ...lar head 1Set Eyepiece Wide field plane scope eyepiece WF10X 18 WF16X 13 1Pair each Objective Achromatic objective for fluorescence device 4X 10X 40X s 100X s oil 1Set Power supply box In AC90 254V Ou...

Page 9: ...lect the different spectrum range Pull the selector 5 out to the limiting position for G green light exciting state Push the selector 5 into the limiting position for B blue light exciting state Put t...

Page 10: ...nsequence Select appropriate lens Immersion oil produces the fluorescence Use special immersion oil The lamp in base is not turned off Turn off the lamp The lenses in collector reflect ambient light t...

Page 11: ...place Disconnect it with the power source and put the dust cover over it after using If it will be not used for a longer time it is the best way to screw the objectives out and place them into the le...

Reviews: