Thrust the excitation type selector (5) into the limiting position “B”. And place a white paper

on the stage. Turn the 10X objective in working. Then turn on the switch of the power supply box

(19) and a light shadow will fall down on the paper. Adjust the focusing rod for condenser (26) and

the focusing knob of the microscope to focus the shadow to be a dot.

4. Preparation of the sample:

The sample can not be too thick. Otherwise the judgment will be affected by the overlapping

cells or impurity.

Thickness of cover glass: 0.17mm;

Seal carrier: no fluorescence and no color;

Immersion oil: Special immersion oil for fluorescence device.

5. Fluorescent microscopy observation:

Turn off the halogen in the base and turn on the mercury lamp.

For the best fluorescence observation effect, it is necessary to close the field diaphragm in

the base or cover the collector with pressboard. Because the lenses in collector possibly

reflect ambient light to disturbance the fluorescent observation.

Push the flashboard B (6) in the middle position to pass the light with specific wavelengths.

According to the different demands, select the different spectrum range:

Pull the selector (5) out to the limiting position for G (green light) exciting state;

Push the selector (5) into the limiting position for B (blue light) exciting state;

Put the selector (5) in the middle position for general biological microscopy observation with

ordinary transmission light;

When the fluorescent microscopy observation is paused, please push the flashboard B (6) to the

left limiting position to stop the stimulating light and reduce the fluorescence quenching.

6. General biological microscopy observation:

Turn off the switch (17) of the power supply box (19) and push the selector (5) to the middle

position “O”. Turn on the switch in the base and adjust the brightness of the halogen lamp by

turning the plate (15) for general biological microscopy observation.

E) Use Scope of the Fluorescent Filter:

Filter

Use

Fluorescent Dye

Lighting

Color

Medicine Examination

Project

B

Blue

EX: 490

DM:

510

BA: 515

P1TC

Pyoktanin Yellow O

Alkalinity

Fluorescein

Acridine Orange

Fluorescein Sodium;

Tetracyclin Sodium;

Green,

Yellow

Orange;

Red

Fluorescence

Immune

Body;

Cell

Nucleus

of

the

Tubercidin Bacillus

Cancer Cell

Red Blood Cell

Protozoan

Venereal Diseases Series

G

Green

EX: 545

DM:

580

BA: 590

Luo Daming;

Luo Daming B200;

Acridine Orange

Orange-red

Fluorescence immune body

*DNA

P.5

Summary of Contents for DW-2002H

Page 1: ...EPI FLUORESCENT MICROSCOPE SERVICE MANUAL Read it carefully before operating and keep it with the microscope...

Page 2: ...oscope It is not only used for fluorescent microscopy observation but also used for general biological microscopy observation 2 No dismantle the equipment Unless you are a microscopic expert or there...

Page 3: ...the microscope otherwise it will damage the circuit and bulb even lead to insecurity 5 Protecting optical parts Never try to contact directly the optical surface of objectives eyepieces and other opt...

Page 4: ...Microscope P 2 C The Principle Function and Main Technical Parameters of the Fluorescence Microscope P 3 D How to Use the Fluorescence Microscope P 4 E Use Scope of the Fluorescent Filter P 5 F How t...

Page 5: ...let Bar 13 Fluorescent Objective 14 Microscope Body 15 Brightness Adjustor 16 Tension Guide Line 17 Switch 18 Accumulator 19 Power Supply Box 20 Ultraviolet Bar 21 Observation Window 22 Epi fluorescen...

Page 6: ...with soft cloth and gasoline promptly Remove the plastic cover away from the Epi fluorescence device 22 and install it on the microscope body 14 with the screw 11 Usually the lamp house in the Epi fl...

Page 7: ...the power supply box 19 and fix it 5 If the objectives are packed in plastic bottles please take them out and spin them in the nosepiece 6 Install the viewing head 3 on the Epi fluorescence device 22...

Page 8: ...lar head 1Set Eyepiece Wide field plane scope eyepiece WF10X 18 WF16X 13 1Pair each Objective Achromatic objective for fluorescence device 4X 10X 40X s 100X s oil 1Set Power supply box In AC90 254V Ou...

Page 9: ...lect the different spectrum range Pull the selector 5 out to the limiting position for G green light exciting state Push the selector 5 into the limiting position for B blue light exciting state Put t...

Page 10: ...nsequence Select appropriate lens Immersion oil produces the fluorescence Use special immersion oil The lamp in base is not turned off Turn off the lamp The lenses in collector reflect ambient light t...

Page 11: ...place Disconnect it with the power source and put the dust cover over it after using If it will be not used for a longer time it is the best way to screw the objectives out and place them into the le...

Reviews: