17
Tubing if there is something small inside, and make the Tubing
dry before using.
• Make sure the Cap Plug is installed correctly.
• Make sure the Seal Ring is assembled.
3. No suction
• Replace the distorted membrane for a new one.
• Make sure the Cap Plug is installed correctly.
• Make sure the Seal Ring is assembled correctly, if damaged,
replace with a new one.
• Make sure all the plastic parts are not damaged by high
temperature sterilizing.
4. Milk overflow
• If milk gets into the Tubing, Damper or Grid, turn off the pump
and unplug from the power source immediately, clean the Tubing,
Damper or Grid according to the cleaning instructions.
• If Damper or Grid is damaged, replace with a new one.
• Make sure your sitting position is appropriate for pumping.
Technical Data
• AC/DC adapter: input 100-240V, 50/60Hz, .2A output 6V 1A.
• Stimulating velocity: 95~105 T/min;
Stimulating intensity: -0.002~-0.019MPa, 9 levels vacuum.
• Expression velocity: 20~65 T/min;
Expression intensity: -0.011~-0.033MPa, 9 levels vacuum.
• Timer: 30 minutes.
• Overall noise:
≤
60dB(A).
• Operating conditions: 5°C to 40°C ,15% to 90% RH,800 hPa to
1060 hPa.
• Storage and shipping conditions:-25°C to 70°C ,15% to