Part II: Turning the System On and Off
Make sure your system is correctly set-up as described in Part 1. It is important to let the SP2
Instrument run for about 20 minutes by itself to warm up and stabilize. Never turn the laser on before
the pump is active, this can damage the instrument, turn on the pump and let it run for some time
before the laser is activated. Once the system is warmed up and working properly, you can replace
the nebulizer cover and secure it using the two back-panel screws.
SP2, and Nebulizer system
Steps for powering up the system and general use.
1.10
SP2 Pre-sampling warm-up
1.
Make sure the Zero Filter Setup is attached (Figure 10).
Figure 10: ZERO filter attached.
2.
Switch on the
Computer
for the SP2.
3.
Switch on the SP2
System Power.
4.
Start the SP2 software by double-clicking on the SP2 icon (Figure 11) on desktop.
Figure 11: SP2 Program launch Icon