DOC-0341 Rev E
2 1
© 2019 DROPLET MEASUREMENT TECHNOLOGIES LLC
5.
The instrument is now ready to be connected to the nebulizer. Connect the nebulizer to the SP2
by attaching the tubing into the “Tee” (Figure
22).
Figure 22: Reconnecting the Nebulizer to the SP2
6.
Note that the Inline filter connected to the “Tee” connector is used to vent the flow and is not
connected to any other components.
7.
Check to be sure that the incandescence and scattering concentrations are near zero. If they are not
near a zero concentration, there is a possibility of a leak in the system. Check the connectors on the
glass fittings of the nebulizer.
Turning off the System
To turn off the system, follow the procedures below.
On the SP2 software’s Control tab, set the laser switch to laser off (Figure
23)
WARNING
:
Laser Power must be turned off before the pump is turned off.
Figure 23: Turning the Laser Off