DOC-0128 Rev F
© 2017 DROPLET MEASUREMENT TECHNOLOGIES
8 2
12.6.1.3
DMT Service Tab: Flow Calib.
Columns A and B each have their own
Flow Calib.
sub-tabsFigure 56 displays Column A’s sub-tab.
The
Flow Calib.
sub-tabs allow users to enter the slopes and intercepts of the equations to convert voltages
to flows. To have the voltage of a column’s transducer displayed in the column’s respective CCN-200
Flows
tab, navigate the Flow Calib. sub-tab for that column. Then enter a slope of 1 and a intercept of 0. This can
be used to display transducer voltages for either the sheath or sample flow, or the absolute pressure. Refer
to section 8.2 for the complete flow calibration procedure.
The
Supply Pump
and
Drain Pump
indicators blink during normal operation when the pumps are on.
Figure 56: DMT Service Tab – Flow Calib. Sub-tab