8.0
Calibration Procedures
For optimum Dual Column CCN performance, periodically calibrate of the Dual Column CCN flow
sensors, pressure transducers and the optical particle counter.
8.1
SS% Calibration
8.1.1
Recommended Calibration Equipment
•
Aerosol Generator System for nebulizing ammonium sulfate particles (DMT Aerosol
Generator AG-100))
•
Differential Mobility Analyzer (TSI3080L or equivalent)
•
CN Particle Counter (TSI3025L or equivalent)
Refer to Figure 23 for the calibration set-up.
The CN Counter is used as a reference instrument. It will count 100% of all the particles that are fed
into it from the DMA. The ratio of the number of particles the CCN could count if they activated can
now be determined.
Figure 23: Calibration Set-Up
Particle
Generator
DMA
CN Counter
DUAL
COLUMN CCN