DOC-0128 Rev F
© 2017 DROPLET MEASUREMENT TECHNOLOGIES
5 2
Problem
Possible Cause
Solutions
“Column Temps Stabilized
Status”
light turns green
occasionally but does not stay
green
.
One or more of the Column
Temperature zones—T1,
T2, or T3—is stabilizing at
more than 0.3° from set
point.
No customer-adjustable solutions. Contact DMT for
further help.
Note: If temperatures are being controlled but
stabilization point is only slightly more than .3
degrees from set point, (< .4), the unit is probably
functioning properly.
Major system failure, computer
not booting, all system values
reading significantly off,
negative values for
temperatures.
Power supply problem
The power supply board and the CCN
control board have LED units for each of the
power supplies; see Figure 19. The supply
voltage is identified on the silkscreen near
the LED and if the LED is not lit, the problem
is most likely in the fuse or power supply.
Table 1 lists the fuses in the CCN, values and
applications. The fuses used are a printed
circuit type, and it is not possible to tell
visually if they have failed. It is necessary to
check the fuses with some type of
continuity device.
Power Supply PCB
Control PCB Column A
Control PCB Column B
Fuse
Value
Application
Fuse
Value
Application
Fuse
Value
Application
F1
5 A
12 DC-DC
converter
F2
3A
OPC power
F2
7A
Top TEC
controller
F2
7A
Top TEC
controller
F3
3A
Computer power
F3
7A
Middle
TEC
controller
F3
7A
Middle
TEC
controller
F4
2A
5V DC-DC
Converter
F4
7A
Bottom TEC
controller
F4
7A
Bottom TEC
controller
F5
2A
Touch screen
power
F5
28 VDC spare
circuit
F5
28 VDC spare
circuit
F7
2A
Nafion heater
F7
2A
Nafion heater
F8
2A
Inlet heater
F8
2A
Inlet heater
F9
2A
OPC heater
F9
2A
OPC heater
Table 1:
Dual Column CCN Fuse Listing