DOC-0425 Rev D
© 2019 DROPLET MEASUREMENT TECHNOLOGIES LLC
1 8
Unpacking, Setup, Installation, and Power Up
1.8
Unpacking
The SP2-XR is shipped in a case specially designed to fit and protect the contents of the shipment.
This case should be saved and used to return the instrument to DMT if service is needed.
The shipment also includes:
•
Calibration Documents
•
Toolkit
•
Alignment Tools
•
Benchtop power supply with regional adapters and aircraft power integration cable
•
Zero count filter
Check all contents for damage immediately upon receipt and notify the carrier and DMT if any
damage is noted.
1.9
Installation
The SP2-XR is shipped with a benchtop power supply (including regional adapters) and an aircraft
power integration cable. If using in a laboratory setting, simply follow section 1.10 Setup to proceed.
If using in a mobile setting and/or on an aircraft, care must be taken in mounting the SP2-XR. See
Appendix G: Mounting Information to ensure that you take care to mount the SP2-XR properly.
In all cases, good practice and care must be exercised to ensure that the aerosol losses in the sampling
system are minimized; including minimizing sample tubing length, ensuring that any inlet that is used
is conductive and basic sampling theory is followed.
1.10
Setup
1. Connect the SP2-XR power cable, either remote power or the provided power supply.
2.
If using the provided power supply, plug in the appropriate region’s adapter. The optional ground
is generally used to protect the instrument in an aircraft.