1.2
Electrical Specifications
Power Requirements:
•
90
–
230 / 50-60 Hz with DMT supplied power supply
•
18-36VDC, (24V Nominal) with direct power connection:
o
Power connector is part number RT00104PNH and it mates
with RT06104SNH
o
Power surge protection is highly recommended
Power Consumption:
25 W typical / 85 W maximum
Backup Battery Capacity
10.24Wh
Front Panel Connections
VGA and HDMI monitor ports, Two Ethernet ports, 2x USB 3.0 ports,
and 2x USB 2.0 ports. 1/8 in. Swagelok® sample connection line, 1/4
in. Swagelok® exhaust line, fan inlet.
1.3
Recommended Maintenance
Weekly Routine Maintenance:
Conduct PSL size check to verify calibration
Monthly Maintenance
(and around field campaigns):
Conducting zero check with high-efficiency filtered air sample.
Annual Maintenance and
Recommended Service
Annual cleaning and calibration at a DMT certified service facility
under normal operating conditions, may need service on a more
frequent basis depending on type of environment where the
instrument is being used
.
1.4
Physical Specifications
Weight:
13 Kg (28.5lbs) for SP2-XR Instrument
Dimensions:
21 cm W x 21 cm H x 42 cm L (Actual) Not including inlet hardware
and fuses.
Shipping Container:
61 cm W x 35.5 cm H x 63.5 cm L (Actual)
32.3 Kg (69 lbs.)