2.2 Flow System
A pump draws on an exhaust jet, pulling flow through the inlet jet and across the laser mode as
illustrated in Figure 6. The inlet jet is an aerodynamically focused assembly with a sample nozzle
of 500 µm diameter and a sheath nozzle of 760 µm diameter. The tip of the sheath jet sits within
1 mm of the edge of the laser mode.
Sample flow is between 1 and 100 sccm, typically 50 sccm. The sheath flow is typically 700 sccm.
Particle velocity depends on sheath flow rate, but is on the order of 50 to 100 m/sec. Particles
are confined to a space that is approximately 10% to 20% of the laser beam diameter, which is
0.5 mm (
e
-2
intensity diameter). This yields a sizing resolution of approximately 2% to 5% of the
particle size.
Figure 6: Schematic Diagram of Flow System