Manual, Ultra High Sensitivity Aerosol Spectrometer (UHSAS)
DOC-0210 Rev E-4
2 1
© 2017 DROPLET MEASUREMENT TECHNOLOGIES
4.0
Using the Instrument
The UHSAS software is designed using LabVIEW, a program from National Instruments that
provides a user-friendly virtual instrument (commonly called a vi) panel for the control, data
display, and data logging of the UHSAS instrument. The sections below provide an overview of
how to use the UHSAS software to start, run, and shut down the instrument. Further details
appear in Part Two of this manual.
4.1 Start-Up
1)
Before starting the instrument, ensure that the following conditions are met:
•
The keyboard and track-pad are connected to the USB port on the front panel of the
instrument
•
The zero filter is in place on the inlet
•
The power switch is in the off position
•
The power cable is connected
2)
Turn on the instrument with the power switch located on the back panel.
3)
Click on the Spectrometer icon on the desktop to bring up the UHSAS window, shown in
Figure 10. Note that the program starts with the
Controls
tab as the default. The
Controls
screen allows the user to set the sample flow and to monitor various parameters.
4)
Allow the instrument to run for 30 minutes so the laser and various heaters stabilize.
5)
On the
Controls
tab, look to see that the
Reference
voltage is 2.5 ± .25 V. If the displayed
reference is less than 1.88 V, clean the laser optics as described in section 7.0.