Manual, Ultra High Sensitivity Aerosol Spectrometer (UHSAS)
DOC-0210 Rev E-4
5 4
© 2017 DROPLET MEASUREMENT TECHNOLOGIES
The
Clear
button clears out old data if users have made adjustments.
The
Run
button initiates relative-gain calibration. The process will continue until the user
presses the
Run
button again or until the instrument has processed the number of particles
stipulated in the
Pts
control.
The
Pts
control sets the number of particles the UHSAS processes before relative-gain
calibration is complete.
The
Align
button, visible only on the
G2:G1
tab, is not used.
The switch in the bottom left of the screen indicates whether the gain-ratio function should be
forced through the (0,0) point. Typically this control should be set to “Allow Non-Zero Offset.”
For step-by-step instructions on relative-gain calibration, see section 5.2.2.
Figure 31: A Relative Gain Sub-Tab