4.2 Check the Flows:
1)
Look at the
Sample Flow
displayed on the left of the screen. The instrument is calibrated at
the factory with a sample flow of 50 sccm, so it is a good idea to set the Sample Flow to this
value during the following performance check. Adjust the flow if necessary, either by using
the “slider” control or by manually entering a number in the set voltage box at the bottom
of the flow indicator.
2)
Look to see that the Sheath flow is about 500-800 sccm (700 sccm is typical at sea level).