Maintenance
7.1 Yearly Maintenance at DMT
DMT recommends returning your instrument to the factory for an annual cleaning and
calibration. This will ensure your instrument is working properly and prolong its lifetime.
7.2 Cleaning the Laser Optics
Under normal operating conditions, the laser optics should not need to be cleaned very often.
However, if the unit is often moved or turned on and off, the optics may need more frequent
cleaning. A good indication of whether the laser optics require cleaning is the
Reference
voltage,
which is displayed on the
Controls
tab. The initial value for the Reference voltage is typically 2.5
V; the factory value for your particular probe can be found on the last page of the data sheet
supplied with the instrument. If this voltage drops by 25% or more, the laser optics will need to
be cleaned.
CAUTION:
The UHSAS uses a high power invisible infrared laser. The optics should only be cleaned by
personal familiar with laser safety precautions. Failure to follow safe practices can lead to eye damage
or permanent blindness.
In no cases should the laser be operated with a cleaning port removed.