Manual, Ultra High Sensitivity Aerosol Spectrometer (UHSAS)
DOC-0210 Rev E-4
3 6
© 2017 DROPLET MEASUREMENT TECHNOLOGIES
To clean the optics, follow the steps below:
1)
After turning off power to the unit, remove the 16 Phillips-head screws holding the
cover on. Lift the cover straight up until it clears the unit.
2)
Open the tubing at the joint in the exhaust line. The exhaust line comes out of the
bottom of the large black optical block (see Figure 20). Opening the line creates
positive pressure in the block and helps keep out contamination.
3)
Lift the power interlock switch at the front corner of the unit and turn on the unit.
4)
Ensure that the laser interlock switch at the rear is off by pressing it downward and
allowing it to spring back to center.
Figure 20: Opening the Exhaust Line
Figure 21: UHSAS Power Switch on Front of Unit