Manual, Ultra High Sensitivity Aerosol Spectrometer (UHSAS)
DOC-0210 Rev E-4
4 0
© 2017 DROPLET MEASUREMENT TECHNOLOGIES
Part II: Software
The UHSAS program is designed in LabVIEW, a software program that provides a user-friendly
virtual instrument panel for the control, data display, and data logging of the UHSAS instrument.
This manual describes version 4.1.0 of the software.
When the UHSAS program is started, five tabs appear at the top of the screen. Clicking on these
tabs will bring up different displays, as follows:
1)
Controls Tab
– Monitors and/or controls variables such as sample flow, sheath flow,
reference and scatter values, ambient pressure and temperature, and laser current and
temperature. The Controls tab is the default screen that appears upon program start-up.
2)
Map Tab
—Allows the user to commit the desired bin map to the instrument. The bin
map determines how the UHSAS categorizes and sizes particles.
3)
Histogram Tab
—Displays a histogram of particles and allows the user to set desired
histogram parameters. These parameters include whether the histogram displays data
linearly or logarithmically, whether the x-axis charts bin number or particle size, and
whether the y-axis displays particle count or concentration. This tab also allows users to
set the sample time.
4)
Configuration Tab
—Allows the user to view and set system parameters. This tab also
allows users to enter a system password which expands access to features and
parameters on the Configuration tab and other tabs.
5)
Calibration Tab
—Assists the user in recalibrating the UHSAS.
The sections below provide details about the different screens available by clicking on the tabs.