8.4 Source Thermocouple (
)
TC
Type
Temp
CJ
CJ Temp
Unit
Mode
(
-/
-keys to change)
Select the type of thermocouple
, , , , , , ,
,
LEFT RIGHT
K J T B R S E N U
L
or
(
-/
-keys to change)
LEFT RIGHT
Internal
: the cold junction temperature is calculated internally
in the calibrator,
Manual
or
: you are able to set the cold junction temperature
manually,
External
or
: for use with the external cold junction compensation unit, supplied with
the calibrator.
(value set keys to change)
If the cold junction is set to manual then you can set the
temperature of the junction here. Valid manual cold junction
temperatures range between -99.9°C and +99.9°C.
(
-/
-keys to change
LEFT RIGHT
)
INCREMENT:
output any value in the range of the
choosen thermocouple
see chapter 8.1 - Source Milliamps.
,
or
RAMP:
Example:
Sourcing 1210.0°C type K thermocouple
(
-/
-keys to change)
,
LEFT RIGHT
Celsius °C
or
or
or
Fahrenheit
,
Kelvin °K,
.
Rankine °R