QBC STAR System Operator’s Manual
6000-300-000
7–10
417
Actuator position error
1. Record the error message displayed on the LCD screen
2. Export the support file (Ref. page 2-8)
3. Place the instrument into Shipping Prep (Ref. page 2-9)
4. Power down the instrument (Ref. page 4-10)
5. Confirm the rotor spins freely by hand
6. If the rotor spins freely
a. Power down the instrument (Ref. page 4-10)
b. Turn the instrument Power on (Ref. page 4-2)
c. If the error persists contact Technical Support
7. If the rotor does not spin freely, contact Technical Support
418
Actuator current error
1. Record the error message displayed on the LCD screen
2. Export the support file (Ref. page 2-8)
3. Place the instrument into Shipping Prep (Ref. page 2-9)
4. Power down the instrument (Ref. page 4-10)
5. Confirm the rotor spins freely by hand
6. If the rotor spins freely
a. Power down the instrument (Ref. page 4-10)
b. Turn the instrument Power on (Ref. page 4-2)
c. If the error persists contact Technical Support
7. If the rotor does not spin freely, contact Technical Support
419
Exhaust temperature sensor error
1. Record the error message displayed on the LCD screen
2. Export the support file (Ref. page 2-8)
3. Power down the instrument (Ref. page 4-10)
4. Tip the instrument up and check for obstructions in the
exhaust port
5. If there are obstructions, clear the obstructions and confirm
the proper setup (Ref. page 2-1)
6. Turn the instrument Power on (Ref. page 4-2)
7. if the error persists contact Technical Support
420
Keypad error
1. Record the error message displayed on the LCD screen
2. Export the support file (Ref. page 2-8)
3. Power down the instrument (Ref. page 4-10)
4. Turn the instrument Power (Ref. page 4-2)
5. if the error persists contact Technical Support
421
Run button error
1. Record the error message displayed on the LCD screen
2. Export the support file (Ref. page 2-8)
3. Power down the instrument (Ref. page 4-10)
4. Turn the instrument Power (Ref. page 4-2)
5. if the error persists contact Technical Support
422
Lid latch release bar error
1. Record the error message displayed on the LCD screen
2. Export the support file (Ref. page 2-8)
3. Power down the instrument (Ref. page 4-10)
4. If the lid is closed, open it manually (Ref. page 4-10)
a. If the lid opens, turn the instrument Power on (Ref. page 4-2)
b. If the error persists contact Technical Support
c. If the lid does not open, contact Technical Support
5. If lid is open, turn the instrument Power on (Ref. page 4-2)
6. If the error persists contact Technical Support
Summary of Contents for QBC STAR
Page 14: ...THIS PAGE INTENTIONALLY LEFT BLANK...
Page 28: ...THIS PAGE INTENTIONALLY LEFT BLANK...
Page 40: ...QBC STAR System Operator s Manual 6000 300 000 4 12 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 50: ...THIS PAGE INTENTIONALLY LEFT BLANK...
Page 72: ...6000 300 000 A 1 8 Appendices 6000 300 000 8 1...
Page 74: ...THIS PAGE INTENTIONALLY LEFT BLANK...
Page 76: ...THIS PAGE INTENTIONALLY LEFT BLANK...
Page 78: ...THIS PAGE INTENTIONALLY LEFT BLANK...
Page 80: ...THIS PAGE INTENTIONALLY LEFT BLANK...