background image

11

Program the (P)WLS909-433 or (P)WLS919-433 Function
Buttons

(P)WLS909-433 and (P)WLS919-433 wireless keys have four programmable function
buttons.  Default functions have been assigned, but you may program other functions if
desired.  After the functions are programmed, when you press and hold one of the four
buttons for one second, the system will execute the programmed function.

For systems using partitions (PC5020/P-8+/PC501X/P832/P832DL only):

 All wireless

keys assigned to Partition 1 will have the four functions programmed in section [61]. All wireless
keys assigned to Partition 2-8 will have the four functions programmed in section [62-68]. For
example, if function button 1 in Section [61] is programmed for Stay arming, then pressing the
first button on wireless keys assigned to Partition 1 will Stay arm Partition 1.

NOTE

: Wireless keys will not work when the partition they are assigned to is being

accessed for zone bypassing or programming.

1. At a system keypad, enter [

][8][Installer’s Code].

2. Enter programming section [804].

3. Enter programming section [61] to [68] for partitions 1 to 8.

4. For each of the 4 function buttons, enter the 2-digit number of the function you want to

select. See the programming worksheets in the back of this manual for a list of function
key options.

5. Record your programming choices in the worksheets in the back of the manual.

6. To exit press [#].

3.5 RF5501-433 PGM Outputs

The RF5501-433 has two on-board open collector PGM outputs.  Each of these can be
individually programmed to:

1. Follow main panel PGM outputs 1 to 14.

NOTE: 

Please refer to your System Installation Manual for available PGM outputs.

NOTE: 

If the RF5501-433 is connected to the PC580/5PC85/PC1555/PC1565/P-48/P-6B

(all versions), PC5010 v1.x, P832 v1.x or the WSS5010 1.0, 2.1, the RF5501 PGM outputs
cannot be programmed to follow main panel PGM outputs 1 or 2.

2. Activate for a programmable amount of time when a signal is received from a wireless

key programmed with output option 31 or 32 (RF5501 PGM pulse) and the PGM
output programming section [70] or [71] is programmed with option 15 (RF5501 PGM
pulse).  The amount of time that the PGM outputs can be programmed to remain on is
programmed in sections [72] and/or [73]. The default activation time is 5 seconds.

3.  Toggle state when a signal is received from a wireless key programmed with option 31

or 32 and the PGM output programming section [70] or [71] is programmed with
option 16 (RF5501 PGM toggle).

3.6 Deleting Wireless Devices

To remove a wireless device from the system, follow the guideline for adding a wireless
device. Program the ESN as [000000]. The wireless device for the zone will be removed.

Now that you have enrolled all the wireless devices, you will need to program the system
to work properly with the devices. See section 4 for more information.

E  N  R  O  L  L    &    P  R  O  G  R  A  M    D  E  V  I  C  E  S

Summary of Contents for RF55O1-433

Page 1: ...433 version 5 O WARNING Please refer to the System Installation Manual for information on limitations regarding product use and function and information on the limitations as to liability of the manuf...

Page 2: ...system may not properly alert occupants of a fire for a number of reasons some of which follow The smoke detectors may have been improperly installed or posi tioned Smoke may not be able to reach the...

Page 3: ...ireless Devices Using Zones 9 3 3 A Note About Electronic Serial Numbers 10 3 4 Enroll Wireless Devices Using Zones 10 3 5 RF5501 433 PGM Outputs 11 3 6 Deleting Wireless Devices 11 Section 4 Other Pr...

Page 4: ...atures Maximum Current Draw 220 mA with back lighting boost on and both PGMs drawing maximum current of 50 mA each Frequency 433 92 mHz Voltage 12V Zones receiver can receive signals from up to 32 wir...

Page 5: ...ference including electrical noise computers televisions electric motors in appliances and heating air conditioning units large metal objects such as heating ducts and plumbing which may shield the an...

Page 6: ...gnment 2nd digit tells the panel which keypad slots are occupied The panel can then generate a fault when a keypad supervisory signal is not present There are eight available slots for keypads RF5501...

Page 7: ...eypad is not connected properly to the Keybus there is a problem with the Keybus wiring run the keypad does not have enough power 1 6 Downloading This product has an integrated RF5501 433 v5 0 receive...

Page 8: ...e see your system s Installation Manual for instructions on programming the keys and a complete list of all the function key options available for your system 2 3 Clock Options The RF5501 433 will dis...

Page 9: ...disabled 2 ON Auxiliary key enabled OFF Auxiliary key disabled 3 ON Panic key enabled OFF Panic key disabled 5 When you are finished press to exit 2 6 Door Chime Options You can program the RF5501 433...

Page 10: ...2 4 Turn one of the following options on by pressing 1 2 3 or 4 1 4 quick beeps default sound 2 Bing Bing tone 3 Ding Dong tone 4 Alarm tone NOTE Ensure that only one of the above options is turned on...

Page 11: ...0 and higher PC1555 MX P 6B MX PC580 585 v2 0 and higher P 48 v2 0 and higher by turning section 015 option 4 off To enable the reporting of openings and closings by identified wireless keys Make sure...

Page 12: ...alnumbersonly WhenconnectingtheRF5501 433toaPC5020 P 8 PC5015v2 x and higher P832DL v2 x and higher PC5010 v2 0 and higher P832 v2 0 and higher PC1555 MX P 6B MX PC580 585orP 48panelenterthe6 digitser...

Page 13: ...ackofthemanual 6 To exit press 3 5 RF5501 433 PGM Outputs The RF5501 433 has two on board open collector PGM outputs Each of these can be individuallyprogrammedto 1 Follow main panel PGM outputs 1 to...

Page 14: ...RF5501 433 via the Keybus after at least one device has been enrolled on the module see section 3 2 Enrolling Wireless Devices To activate module supervision after you enroll the first device s 1 Exit...

Page 15: ...ption disabled Wireless Supervisory Window Each wireless zone will send a supervisory signal every 64 minutes If the receiver does not receive a signal within the time programmed for the Wireless Supe...

Page 16: ...or more information 5 RF jamming detection is now enabled To exit Installer programming press NOTE A jamming condition will be indicated as RF Jam Zone if section 93 is pro grammed regardless of the R...

Page 17: ...or from its backplate wait for 5 seconds then reattach the detector to its backplate Or hold a magnet near the raised line on the outer rim Then remove the magnet WLS907 433 WLS925L 433 Open and close...

Page 18: ...r Moving the RF5501 433 higher will usually improve the reception If you move the RF5501 433 repeat the tests described in sections 5 1 and 5 2 on all the wireless devices Continue to test the devices...

Page 19: ...e programmed for the Wireless Supervisory Window it will generate a zone fault Tamper Switches Removing the RF5501 433 from its mounting location will cause a general system tamper Wireless Zone Low B...

Page 20: ...le placement test I get no result or Bad results Check the following see sections 5 1 and 5 2 for more information on testing devices Verify that you are testing the correct zone Verify that the corre...

Page 21: ..._ _ _ _ _ _ _ I 1 Local Clock Display Enabled Display Disabled ON I _ _ _ _ _ _ _ _ I 2 Local Clock Displays AM PM Displays 24 hour Time OFF I _ _ _ _ _ _ _ _ I 3 Open Zones Override Clock Display Do...

Page 22: ...________I I________I I________I I________I 16 I_____________________ _____________________________________________I I________I I________I I________I I________I 17 I_____________________ ______________...

Page 23: ..._______I I________I I________I 48 I_____________________ _____________________________________________I I________I I________I I________I I________I 49 I_____________________ __________________________...

Page 24: ...l_____l_____l_____l_____l_____l_____l 13 Zone 13 l_____l_____l_____l_____l_____l_____l 14 Zone 14 l_____l_____l_____l_____l_____l_____l 15 Zone 15 l_____l_____l_____l_____l_____l_____l 16 Zone 16 l___...

Page 25: ...t p i r c s e D y e K n o i t p i r c s e D y e K n o i t p i r c s e D y e K y r t n E y r t n E y r t n E y r t n E y r t n E n o i t p i r c s e D y e K n o i t p i r c s e D y e K n o i t p i r c...

Page 26: ...30 l____l____l Partition 6 Wireless Key Options 66 Function Key 1 03 l____l____l Function Key 3 27 l____l____l Function Key 2 04 l____l____l Function Key 4 30 l____l____l Partition 7 Wireless Key Opt...

Page 27: ...Default 00 Default 00 Default 00 PGM1 Output Activation Time Minutes l_ _______ l________ l Default05 Default05 Default05 Default05 Default05 PGM1 Output Activation Time Seconds l_ _______ l________ l...

Page 28: ...l________l Option 1 Zone 09 Supervision enabled Disabled l________l Option 2 Zone 10 Supervision enabled Disabled l________l Option 3 Zone 11 Supervision enabled Disabled l________l Option 4 Zone 12 S...

Page 29: ...ion 7 Zone 31 Supervision enabled Disabled l________l Option 8 Zone 32 Supervision enabled Disabled 90 Other Options Default Option ON OFF OFF l________l 1 5 For Future Use ON l________l 6 Keypad Zone...

Page 30: ...ns or windows Rapid air movement around the detector may prevent smoke from entering the unit Do not locate detectors in areas of high humidity Do not locate detectors in areas where the temperature r...

Page 31: ...ned for replacement under warranty will be credited or replaced at DSC s option Products not covered by this warranty or otherwise out of warranty due to age misuse or damage shall be evaluated and a...

Page 32: ...guarantee that interference will not occur in a particular installation If this equipment does cause interference to television or radio reception which can be determined by turning the equipment off...

Reviews: