25 |
P a g e
014010-001 Rev 01
©2019 Data Sciences International.
1)
Locate the blue down-
arrow on the generator’s control panel
a
nd press it eight times to bring up the “Service
Mode” menu.
(Fig 33)
Figure 33
2)
Press the OK button. This moves
the lung’s
actuators into a safe position for removing the lung.
3)
Power down the Individual Cigarette Puff Generator (by turning its main power switch to OFF) and remove the
power cord from the power socket.
4)
Unscrew the screws around the top of the unit to remove the top plate, revealing the inside of the unit. (Fig
34)
Figure 34