Operation 
Prior to completing the electrical connections outlined below, ensure that the work place is free 
of conductive objects that could pose electrical hazards to the user should the objects make 
contact with the high voltage output screw terminals of the VF-500 amplifier. 

 

Connect the load to the VF-500 via the OUTPUT SMA connector 

 

Connect line power to the VF-500 through the receptacle on the back panel 

 

Connect the input signal to the VF-500’s INPUT BNC connector 

 

Adjust the input signal as needed to achieve the desired output from the piezoelectric load 

 
Circuitry Protection 
The VF-500 circuitry incorporates over-current, over-temperature and short-circuits protection. 
 
When the protection circuitry is enabled as a result of excessive operating conditions or 
temperature, the amplifier will disconnect the output temporarily and will attempt to restart.  
During restart, the amplifier will recheck the operating conditions and will go into protection 
mode again if fault conditions exist.  This protection thus protects both the amplifier and the 
load.  
 
The VF-500 is equipped with temperature sensors that will protect the circuitry if the internal 
temperature of the unit exceeds the safe operating condition. The over-temperature condition 
triggers at approximately 70C.  Although the signal from the input BNC will continue to be fed 
into the amplifier and the amplifier will remain connected to the load, the output signal will be 
restricted to a small voltage oscillation (20-30 mV) around earth ground until the sensor indicates 
that the unit has reached approximately 50C.  There is no indication to communicate the over-
temperature condition to the operator. 
 

If the equipment is used in a manner not specified by DSM, the protection provided by the 
equipment may be impaired. 

 
Recommended Environmental Conditions for Operation 
The VF-500 is designed to operate in non-hazardous environments.  Any deviation from the 
specified environment must be approved by DSM and may result in the voiding of any warranty.  
Any deviation not approved by DSM will be grounds for voiding of any warranty. 
 

 

Environmental Temperature: 

 

 

 

0-50C 

 

Percent Relative Humidity: 

 

 

 

0-80% non condensing 

 

Operating Supply Voltage:  

 

 

 

Refer to label at power receptacle 

 
 
 
 
 
 
 
 

Summary of Contents for VF-500

Page 1: ...Parkway Court Suite 160 Franklin TN 37064 Tel 615 595 6665 Fax 615 595 6610 VF 500 VOLTAGE FOLLOWER LINEAR AMPLIFIER FOR PIEZOELECTRIC CAPACITIVE LOADS USER MANUAL DYNAMIC STRUCTURES AND MATERIALS LL...

Page 2: ...ended AC input connector configured for the available line voltage OPERATINGAND SERVICES PRECAUTIONS Operate the amplifier in an environment free of flammable gases or fumes To ensure that its safety...

Page 3: ...plifier is designed to drive capacitive loads only Do not use this amplifier to drive resistive loads The amplifier will only work properly when a capacitive load is connected to the output Never conn...

Page 4: ...ximum operating range of 30 to 150V piezo devices which have two lead wires output voltage and ground 2 Low voltage piezoelectric materials rated for a maximum operating range of 30 to 200V piezo devi...

Page 5: ...an output voltage range corresponding to the specifications of the customer s load one of the two types discussed in the Intended Use section Attach a standard load i e a piezoelectric actuator with...

Page 6: ...in if fault conditions exist This protection thus protects both the amplifier and the load The VF 500 is equipped with temperature sensors that will protect the circuitry if the internal temperature o...

Page 7: ...or its designated repair facility Freight charges to return equipment or software to DSM are the customer s responsibility DSM will pay freight charges to return the equipment to the customer Warranty...

Page 8: ...2 Consequential or incidental damages such as property damage and incidental expenses 3 Conditions caused by external factors such as abuse misuse inadequate power supply or acts of God Owing to the w...

Reviews: