Communication Features
Communication Method
The TDAS G5 DAS supports the industry-standard Ethernet 10/100BaseT/Tx communi-
cation method. Communication is enabled after the power-up sequence has completed
(~20 seconds). (See Appendix D for the network parameters of your equipment.)
LED Indicators
The TDAS G5 DAS has five LEDs (red/yellow/green) which provide ongoing status
information. LED behavior is summarized in the table below.
COM: Provides communication status with the docking station
and DAS.
CAL:
Provides calibration status.
PWR: Provides status on power to the DAS.
ARM: Provides arm, start record, and event status.
STAT: Provides general status.
TDAS G5
DAS LED
State
Description
Computer communicating with docking station
No active communication
All calibrations passed
Calibration in progress
Calibration fault
Power on and OK
Power warning (9.6
–
10.2 V)
Power critical (<9.6 V)
System armed and waiting for start record or event signal
Received start record signal; waiting for event signal
TDAS G5 DAS no longer armed or no longer collecting data or
received event signal
No faults detected
Received event signal
Fault occurred, either:
- Low power input
- Docking station fault