Page 17
Section 3 – Installation
Dukane Manual Part No. 403-591-01
Connecting Cables
Basic Connections
Complete these basic connections for the standard con-
figuration as shown below:
•
AC Line Input
•
System Control Inputs/Status Outputs
•
Ultrasound Output
•
Grounding
Details about the various system connectors and their pin
assignments are covered in the next section.
1. Wire the AC line connector, and attach it to the gen-
erator’s power inlet connector, matching the power
source line, ground, and neutral with the generator’s
line, ground, and neutral connector pins -
A
in Figure
3–4. (See Figure 2-1 also.)
2. If used, wire the user-supplied automation system con-
trol inputs/status outputs to the
P1 SYSTEM I/O
con-
nector, and attach it to the
P1 SYSTEM I/O
port -
B
in Figure 3-4.
3. Attach a high–voltage coaxial ultrasound cable (from
the ultrasonic probe) to J1, the ultrasound output con-
nector -
C
in Figure 3-2.
4. Connect the included ground wire from the grounding
stud,
D
in Figure 3-4, to earth ground.
Figure 3-2
Generator Front View
NOTE
AC Power Inlet
The unit does not include a power switch,
and is powered ON whenever the AC line
power is live. The unit can be switched ON/
OFF with a user-supplied AC circuit breaker
wired to the AC power inlet connection.
NOTE
Connecting Cables
Two-piece pluggable terminal block connec-
tors are used for the System I/O connections
and the AC Power Inlet connections.
This type of connector allows the wiring to
be attached to the spring loaded terminal
connector, which plugs into the mating con-
nector on the
iQ
Auto system front panel.
In the event a field replacement unit is re
-
quired, the screw terminal connectors with
the wires can be easily detached and then
plugged into the replacement unit.
See CAUTION
note on Page 21 .
Grounding Stud
D
P1 System I/O Connector
B
Power Inlet
A
U/S Connector
C