4
Safety Precautions
Before using your Ultrasonic Cleaner, please read and thoroughly understand
these safety precautions. Failure to follow them may result in serious personal injury or
property damage
.
To avoid electrical shock:
•
Unplug from power source before filling or emptying the tank.
•
Keep the control panel and the area around the cleaner clean and dry -- wipe up
solution which spills over the tank brim. Water and high voltage can cause electrical
shock
.
•
Do not operate the cleaner without proper grounding
.
•
Do not remove the grounding prong on the line cord plug
.
•
Do not disassemble your cleaner -- high voltage inside the cleaner is dangerous
.
•
Do not immerse the cleaner in water.
To prevent personal and/or property damage:
•
Operate the cleaner with a vented cover or no cover.
•
Use water-based solutions
.
•
Do not ever use alcohol, gasoline, or flammable solutions
.
Doing so could cause a fire
or explosion and will void your warranty
.
Use only water-based solutions.
•
Do not ever use mineral acids. These could damage the tank.
•
Do not touch the stainless steel tank or cleaning solution -- they may be hot.
•
Do not allow fluid temperature to exceed 70°C (160°F).
•
Do not place your fingers or hands into the tank while the cleaner is operating
.
Doing so
may cause discomfort and possible skin irritation
.
Avoid contact with solutions and
provide adequate ventilation
.
•
Do not use solutions containing chlorine bleach
.
To prevent damage to the cleaner:
•
Change your solution regularly.
•
Do not cover vents on the housing.
•
Do not operate the cleaner dry
.
•
Do not place parts or containers directly on the bottom of the cleaning tank; use a tray
or wire to suspend items. Failure to comply may cause transducer damage and will
void your warranty.
•
Do not allow the solution to drop more than 1½ inch from top of tank with heat or
ultrasonics on
.
Failure to comply may cause transducer and/or heater damage and will
void your warranty
. The fill line
is 1” from top of the tank.