GENERAL SPECIFICATION
Max. voltage between terminals and
earth ground
CAT II 600V
Fuse Protection
F 200mA / 250V
Power
9V battery, NEDA 1604 or 6F22
Display
LCD, 1999 counts, updates 2-3 seconds
Measuring Method
Dual-slope integration A/D converter
Overrange Indication
Only figure “1” on the display
Polarity Indication
“-” displayed for negative polarity
Operating Environment
0°C to 40°C
Storage Temperature
-10°C to 50°C
Low battery indication
“ ” appears on the display
Size
138mm x 69mm x 31mm
Weight
Approx 170g
Accuracy is specified for a period of one year after calibration and at 18 to 28°C (64°F
to 82°F) with relative humidity to 80%.
DC VOLTAGE
Range
Resolution
Accuracy
200mV
100µV
± 0.5% of rdg ± 2 digits
2V
1mV
± 0.5% of rdg ± 2 digits
20V
10mV
± 0.5% of rdg ± 2 digits
200V
100mV
± 0.5% of rdg ± 2 digits
600V
1V
± 0.8% of rdg ± 2 digits
Overload protection: 250V rms. For 200mV range and 600V dc or rms, ac for other
ranges.
DC CURRENT
Range
Resolution
Accuracy
200µA
0.1µA
± 1% of rdg ± 2 digits
2mA
1µA
± 1% of rdg ± 2 digits
20mA
10µA
± 1% of rdg ± 2 digits
200mA
100µA
± 1.5% of rdg ± 2 digits
10A
10mA
± 3% of rdg ± 2 digits
Overload protection: F 200mA/250V fuse (10A range unfused).