plug). Inform the technician responsible.
11. MAINTENANCE INTERVALS – USER / TECHNICIAN
Maintenance due to be carried out
Chapter
Interval
• Setting the pressure reducer
12.2
Once a year
• Draining off condensation water
12.3
Compressors
without
dry-air system
Once a month
At a high degree of humidity
Once a day
Compressors
with
dry-air system
Check every six months,
draining if necessary
• Safety-valve check
12.4
Every six months
• Filter replacement
12.5
Once a year
12. MAINTENANCE
Repair work exceeding normal
maintenance operations may only be
carried out by a suitably qualified
engineer or our customer service
staff.
Only use spare parts licensed by the
manufacturer as well as accessories
destined for the purpose.
Prior to performing any maintenance
or repair work, be absolutely sure to
switch off the compressor first and
disconnect from mains (pull out
mains plug).
To ensure proper functioning of the com-
pressor, be sure to regularly carry out main-
tenance as described in Sections 12.1 to 12.5.
12.1 Pressure Reducer
This part is sold as a special accessory
item and is therefore not included in the
scope of delivery.
The pressure reducer (40) serves to control the
flow pressure in alignment with the desired
working pressure. The pressure reducer is
connected to the pressure switch via an
express coupling (20).
40
15
20