5
2. PRODUCT INFORMATION
2.1 Use for the Intended Purpose
The compressor is intended to be used for
generating compressed air required for
operating dental units of equipment or for
similar applications.
Installation in medical care facilities:
In designing and constructing the
compressor, allowance has been made for the
requirements of medical products where
applicable. Accordingly, the unit can be used
for installation in medical care facilities.
If the unit is installed in medical care facilities,
the requirements stipulated in Directive 93/42
EEC IEC 601-1 as well as the relevant norms
must be observed as applied to installation
and assembly.
2.2 Use other than that for the
Intended Purpose
The compressed air produced by
the compressor is unsuitable for
operating breathing equipment or
similar facilities without additional
filters required for the operating
area.
• The compressors are designed to be
operated in dry, ventilated rooms, ambient
tempe10 to +40 °C.
• Do not expose the compressor to rain. The
machine must not be operated in a damp or
wet environment. Use is also prohibited in
proximity to gases or combustible liquids.
• Prior to installing the compressor in medical
facilities, it must be ensured that the avail-
able medium complies with the requirements
stipulated for the relevant purpose in each
individual case. Observe the particulars
given in Chapter 4. "Technical Data".
When installing, classification and
conformity rating must be carried out by the
manufacturer of the ultimate product.
• Any other use or use beyond what is
specified is deemed to be not for the
intended purpose. The manufacturer
accepts no liability for damage resulting
therefrom. All risk is borne solely by the
operator/user.
2.3 Product Description
The compressor generates an oil-free, dry and
filtered compressed air required for operating
units of dental equipment.
3. SCOPE OF DELIVERY
Compressor with / without
dry-air system
Instructions for assembly
and use .................................... 9000-610-38/30
Accessories incl. pressure hose
ø 8x3x14, length 1 m .................... 5410-002-00
Only for compressors with dry-air system:
Collecting trough .......................... 3413-001-00
3.1 Special Accessories
The parts listed below are
not
included in the
scope of delivery.
Please order as required!
5150, 5151, 5170, 5171,
5250, 5251, 5270, 5271 ............... 5110-500-00
4151, 4251, 5351, 5451 ............... 4251-500-00
Pressure reducer ......................... 6040-992-00
3.2 Consumables
Filter cartridge suitable for compressor model:
Primo, Duo, Duo-Tandem,
Trio, Quattro, Quattro Tandem ..... 0832-982-00
Tornado ........................................ 5430-982-00
Fine-mesh filter for
dry-air system............................... 1610-121-00
Sterilisation filter for
dry-air System .............................. 1640-981-00