Error
Possible cause
Remedy
Compressor starts up, no dis-
play on the operating panel
Operating panel defective
❯
Disconnect the mains plug
and inform a service techni-
cian.
Compressor will not start
No display on the operating
panel
No mains voltage
❯
Check the main power switch,
mains fuse and mains voltage,
inform an electrician if neces-
sary.
Fault button flashes (if compres-
sor is equipped with 2 units)
Emergency mode possible
❯
Activate emergency mode:
Press the fault button, (see
"13.8 Emergency mode").
Compressor runs with 1 unit.
❯
Inform a service technician
Fault button lit up
Compressor defective
❯
Disconnect the mains plug
and inform a service techni-
cian.
LED on the filter replacement
button lights up
Filter replacement required
❯
Change all filters
The fault button flashes
Compressor unit defective
❯
Activate emergency mode:
press the fault button. Com-
pressor runs with 1 unit
❯
Inform a Service Technician.
Vent valve defective
❯
Activate emergency mode
and inform a service techni-
cian.
Fault button lit up
Unit is defective
❯
Unplug the mains plug.
❯
Inform a service technician.
Troubleshooting
38
4252100034L02 2105V002